Vacuum system upgrade. Vacuum probe installed Spectrometer sealed – vacuum improved from 300 Pa to 7 Pa. We have to change a hoose for further improvement . Efect on spectra is significant. Electronics upgrade. - PowerPoint PPT Presentation

Transcript of Vacuum system upgrade

115.1.204 V. Vorobel, SuperNEMO meeting, Aussois

Vacuum system upgradeVacuum probe installedSpectrometer sealed – vacuum improved from 300 Pa to 7 Pa.We have to change a hoose for further improvement.

Efect on spectra is significant

215.1.204 V. Vorobel, SuperNEMO meeting, Aussois

Electronics upgradePC controlled disriminator CAEN installed in VME – simplification of electronics, well defined and stable threshold, no possible interference between crates

Independent HV sources for different PMTs – to equilibrate differences in gains. We are waiting for delivery of VME PC controlled HV modules.

Cables fixation

Waiting for new PC controled HV modules and repaired VME doesn‘t prevent measurement.

315.1.204 V. Vorobel, SuperNEMO meeting, Aussois

Magnetic history influence

Does the electron beam energy depend on the previously set energy due to the magnetic hysteresis?1) If we keep the rule – first go up to magnetic saturation then go down to the desired energy2) If we keep the rule – don‘t keep 1) if the desired energy < the original energy

415.1.204 V. Vorobel, SuperNEMO meeting, Aussois

Swapping measurement

Does the peak position depend on the measurement position in the Black-Box due to magnetic field variations?

515.1.204 V. Vorobel, SuperNEMO meeting, Aussois

Magnetic shielding upgrade- 4 m-metal cylinders internal f133mm,length 305mm, thickness 1mm for PMT shielding

instead of the low C steel cylinders which were used up to now- plan to make a m-metal box (thickness 1mm) to cover the spectrometer completly

615.1.204 V. Vorobel, SuperNEMO meeting, Aussois

Magnetic shielding measurement- Magnetic sensor placed instead of PMT in measurement possition 1.- B-field measured in 3 directions in 3 shielding configurations (no shield, low C steel, m-metal)

for currents in the coil corresponding to the electron beam energies 0-1600keV, saturation.

0 500 1000 1500 2000 2500 3000 3500-50

0

50

100

Series1Series3Series5

0 500 1000 1500 2000 2500 3000 3500

-60-50-40-30-20-10

0

Series1Series3Series5

0 500 1000 1500 2000 2500 3000 3500

-20-15-10-505

Series1Series3Series5

7

PMT HV optimization

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

8

DE detector setting optimization

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

9

DE detector setting optimization

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

10

DE detector setting optimization

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

11

DE detector setting optimization

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

- PMT HV 1300 V- DE HV 1200 V- Vacuum 6.4 Pa- E=1000 keV- R=13.6 %

12

Oscilloscope snap

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

DE left

DE right

scint. block

13

Observed spectra

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

- PMT HV 1300 V- DE HV 1200 V- DE threshold 15 mV- Vacuum 7 Pa

14

Conclusions

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

15

Backup

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

16

Magnetic shielding measurement

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

17

Electronics view

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

18

Testing station setup

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

19

Routine measurement setup view

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

20

Oscilloscope snap 1

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

DE left

DE right

scint. block

21

Oscilloscope snap 2

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

DE left

DE right

scint. block

22

Oscilloscope snap 3

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

DE left

DE right

scint. block

23

Oscilloscope snap 4

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

DE left

DE right

scint. block

24

Oscilloscope snap 5

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

DE left

DE right

scint. block

25

Oscilloscope snap 6

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

DE left

DE right

scint. block

26

Oscilloscope snap 7

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

DE left

DE right

scint. block

27

Oscilloscope snap 8

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

DE left

DE right

scint. block

28

Oscilloscope snap 9

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

DE left

DE right

scint. block

29

Oscilloscope snap 10

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

DE left

DE right

scint. block

30

Oscilloscope snap 11

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

DE left

DE right

scint. block

31

Oscilloscope snap 12

15.1.204 V. Vorobel, SuperNEMO meeting, Aussois

DE left

DE right