Post on 19-Dec-2015
Uncertainty considerations for the calibration of transfer standard radiation thermometers
Graham Machin, NPL
Abstract
Three broad areas to consider – when formulating Appendix C entry 1.4 “Standard Radiation Thermometers”
ITS-90 scale realisation (fixed point and reference
thermometer)
Uncertainties arising from the radiance source (blackbody)
Uncertainties arising from the transfer radiation thermometer
-------------------------------------------------------------------------------
Finally a few remarks about … MRA Appendix C entries
Introduction
Concerned only with providing cost effective calibration service –
NOT absolute best can do – but near best measurement capability
ITS-90 above the silver point only, according to the formal definition
Measurement equation for scale realisation uncertainties – that
given in the ITS-90 text – two general contributions
1) the defining fixed point blackbody
2) the reference thermometer
ITS-90 realisation uncertainties – fixed point realisation
Following factors to be considered: Intrinsic repeatability of freezes – type A Impurities – departures from 100% purity Departure from emissivity =1 Temperature drop across cavity bottom – due to energy loss
through the aperture
a) all type B
b) taken together for well designed source <10 mK (k=1)
ITS-90 realisation uncertainties – reference radiation thermometer
Spectral characterisation
Non-linearity and gain ratios
Secular effects (drift)
Radiance transfer effects (characterised [for e.g.] by SSE)
Spectral characterisation uncertainties - 1
Spectral responsivity – usually monochromator – U generally type B
Monochromator uncertainties - wavelength stability/accuracy
- repeatability scan to scan (>3 scans then type A)- resolution+stray light
Reference thermometer uncertainties- secular stability of interference filters (stochastic)- out-of-band transmission- temperature coefficient of filters- alignment
Spectral characterisation uncertainties - 2
Other issues – all type B
a) calculation of effective wavelength
b) use mean effective wavelength at gold point – what uncertainty does this introduce
c) detector responsivity uncertainty over filter pass-band
Wavelength uncertainties characterised by:
u=(T90-Tref)(T90/Tref)(/)(1/3)
Effective wavelength of 650 nm and 906 nm filters since 1994
Effective wavelength at 650 nm
650.60
650.70
650.80
650.90
651.00
651.10
651.20
651.30
651.40
0 2000 4000
Radiance temperature /K
Eff
ecti
ve w
avel
eng
th /n
m
1994
1997
1999
Effective wavelength at 906 nm
906.10
906.20
906.30
906.40
906.50
906.60
906.70
906.80
906.90
907.00
907.10
0 2000 4000
Radiance temperature /KE
ffec
tive
wav
elen
gth
/nm
1994
1997
1999
Reference photocurrent, non-linearity, gain ratios
Reference photocurrent – from fixed point
u = (T902 /c2) (IRef/ IRef): typically ~1e-4 (type A)
Non-linearity – detector and electronics on one gain setting Non-linearity – inter-gain setting (type B)
SSE – formal uncertainty estimate
SSE – two approaches, formal or pragmatic
Formal – calculate effective target diameters for reference source and blackbody target, apply SSE correction
– combine (quadrature) uncertainties of each SSE estimate the
type A uncertainty
u = (T902 /c2) (SSE)
SSE – pragmatic uncertainty estimate and inter-calibration drift
Pragmatic (for low SSE systems) – calibrate at diameter X mm use up to target diameter Y mm - SSE=SSE(Y) – SSE(X)
Same equation as previous slide but type B
------------------------------------------------------------------------------------ Secular drift – stability of reference thermometer (e.g.
electronics) - type B – largest component up to 2000 °C – reduced by more frequent fixed pt. calibrations
u=(T90/Tref )2 Tdrift (1/ 3)
Typical reference thermometer uncertainty in scale realisation at 650 nm
Reference thermometer uncertainty
0.00.10.20.30.40.50.60.7
1000 1500 2000 2500 3000Radiance temperature/°C
Un
cert
ain
ties
/°C
wvlgth
ref
N/L
SSE
drift
u(k=1)
Second level MRA CMC entry 1.4 calibrations
Above described top-level calibration
Below describe some uncertainty considerations for “Standard Radiation Thermometers” – laboratories who do not hold a primary calibrated RT but a transfer thermometer calibrated elsewhere IS their standard RT
Limited to calibration of RT by comparison using a transfer radiance source
Uncertainties arising from the radiance source
Assume blackbody or quasi-blackbody (emissivity >0.99)
Factors to be considered:
Stability during test – type A Uniformity across test area – type B - see later Wavelength dependence (see later)
Uncertainties from transfer thermometer - I
Repeatability of reference thermometer output at test temperature (type A)
Repeatability of transfer thermometer output at test temperature (type A)
Thermometer resolution – type B
Uncertainties from transfer thermometer - II
Uncertainties associated with corrections for RH and internal thermometer temperature – type B
Standard uncertainty of any ancillary equipment used – e.g. DVM
Uncertainty arising from SSE – strictly negligible as reference thermometer and transfer thermometer are viewing same aperture
- when used as transfer standard due care must be taken to equalise the aperture and uniformity of transfer sources – otherwise large uncertainties can accrue.
Uncertainties from transfer thermometer - III
Mismatch in wavelength between reference and transfer thermometers mod(((s - t)/c2).T2
90 .(1-).(1/3)) – type B
(assume ~1)
Mismatch in target sizes – type B (zero for uniform source)
- otherwise (T/d).s.(1/3) i.e. radiance gradient x nominal target size – (arbitrary >98% of signal taken to be target size s)
Short term repeatability (alignment) – type A if low order fit used
- type B if repeat point differences used
Summary of uncertainty analysis
To arrive at the uncertainty in the calibration of a transfer thermometer requires clear knowledge of:
Scale realisation uncertainty – top level 1.4 cmc entry Transfer source uncertainty plus…. that associated with both the calibration of and intrinsic to the
transfer thermometer – secondary level 1.4 cmc entry
Worked example
Source of uncertainty Value Distrib-ution
Divisor Conversion factor
u/°C Comments T=2200 °C
Reference thermometer 0.10 N 1 1 0.10 Transfer thermometer 0.20 N 1 1 0.20 Thermometer resolution 0.10 R 1.73 1 0.06 Scale realisation 0.30 R 1 1 0.30 Corrections for RH 0.00 R 1.73 1 0.00 Corrections for ambient 0.00 R 1.73 1 0.00 DVM uncertainty 0.00 R 1.73 0.001 0.00 Units V: u=0.6V Uncertainty due to SSE 0.00 R 1.73 425.11 0.00 (same target) Wavelength mismatch 0.15 R 1.73 1 0.09 650 nm, 1000 nm,
=0.999 Target size mis-match 0.10 R 1.73 3 0.17 3 mm target for TT Short term repeatability 0.20 N 1 1 0.20 Low order fit Combined U /°C 0.47 Expanded U /°C k=2 0.94
Appendix C of MRA - I
What values are to be put in the Appendix C?
Primary scale realisation (reference thermometer) uncertainties?
Transfer thermometer calibration uncertainties?
Appendix C of MRA - II
Technical supplement T7 states “The calibration and measurement capabilities … are those ordinarily available to the customers of an institute through its calibration and measurement services; they are sometimes referred to as best measurement capabilities”
Similar statement in the MRA Glossary –
Calibration and measurement capability “the highest level of calibration or measurement normally offered to clients, expressed in terms of a confidence level of 95%, sometimes referred to as best measurement capability”
Appendix C of MRA – conclusions
From these statements it is reasonable to conclude that:
Appendix C entry not intended to be the best we can attain in near ideal circumstances
Nor is it to include one-off special calibrations
- rather: routine calibrations readily achievable following set procedures - calibrations of good (near-ideal) but real instruments- calibrations for which we would issue a certificate (see T7)