Specifications for the CARESTREAM DRX Plus Detectors
Confidential
Important
• Qualified service personnel must perform these procedures.
• See the Carestream Service Portal for possible updates before you begin.
• When doing the procedures in this document, you must use safe work practices and wear the correct personal protective equipment (for example, safety eyewear) according to your company’s standard operating procedures.
• When servicing this equipment, be sure to read and follow the Manufacturer’s Safety Data Sheets for any chemical, cleaner, solvent, paint, lubricant, adhesive, or replacement battery to understand proper use and disposal precautions.
All rights reserved. No part of this manual may be reproduced or copied in any form by any mean graphic, electronic, or mechanical, including photocopying, typing, or information retrieval systems without written permission.
AC3451 | 2016-04-29 1–1
1 DRX Plus 3543 Detector
Detector
Imaging Specifications
Detector Type Amorphous Silicon on glass—no tiling
Conversion Screen Attached Gd202 S:Tb
Pixel Size (μm) Resolution Limit (cyc/mm)
139 x 1393.6
Usable Pixel Area (cm) (inch)(pixels)
35.4 x 42.514 x 16.72544 x 3056
MTF (%) RQA-3 Beam RQA-5 Beam
0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical
9983614228191310
987954362316107
DQE (%) ±2 sigma < 10 %
IEC 62220-1-1 Edition 1.0, 2015
RQA-3 Beam RQA-5 Beam
0.97 μGy
3.1μGy
9.92 μGy
0.97 μGy
3.1 μGy
9.92 μGy
0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical
3532272114953
38343025191385
373430252015106
3430251813953
35312520151074
35302520151174
Energy Range 40–150 kVp
DRX Plus 3543 Detector
1–2 AC3451 | 2016-04-29
Scan Method Progressive
A/D Conversion 16 bits
Imaging Specifications
Supported Data Interfaces
Wireless 802.11 A, or N, or G
Tethered 10/100/1000 Ethernet
Environmental
Shock High shock tolerance
Temperature Range
Operating +15 to +30° C (+59 to +86° F)
Ambient storage and shipping -10 to +66° C (+14 to +150° F)
Relative Humidity, Non-Condensing
Operating 10 to 86 %
Storage 10 to 86 %
Ingress Protection IP57
Mechanical
Size (cm) 35 x 43 Cassette (ISO 4090)38.35 x 45.95 x 1.47 cm
Weight 3.06 kg (6.75 lb)
Housing Material Aluminum
Sensor Protection Material Carbon fiber and aluminum plate
Load Limit Applied to a single 4 cm (1.6 in.) point: 114 kg (250 lb)Distributed evenly over the detector area: 170 kg (375 lb)
Detector Battery
Technology • Lithium-polymer Technology
• “Smart” battery technology prevents overcharge
Size 21.0 x 15.0 x 0.67 cm
DRX Plus 3543 Detector
AC3451 | 2016-04-29 1–3
Wireless System
Weight 0.4 kg (12.4 oz.)
Voltage / Energy 14.6 to 14.8V DC, 2.1 to 2.5Ah capacity
Battery "Hot Swap" Capability Yes
Charge Capability • 340 maximum images per charge in Direct Connect Mode.
• 14 days per charge in undisturbed Sleep Mode.
Expected Life 500 charge / discharge cycles results in ~80% full charge energy
Medical Safety IEC 62133 —Safety requirements for secondary cells and batter-ies containing alkaline or other non-acid electrolytes
Electromagnetic Compatibility IEC 60601-1-2 Ed. 2.1-Medical Electrical Equipment Electromag-netic Compatibility Requirements and tests, including CISPR 11:1999 + A2:02 emissions Group 1, Class A)
Detector Battery
Technical Specifications
Network Protocol TCP/IP, IPv4, IPv6
Network Type Isolated private wireless LAN (WLAN)
Wireless Protocol 802.11 A, or N, or G
Antenna:Frequency Band Available Channels (fixed at installation)
5 and 2.4 GHz 1, 5, 9, 13, 36, 40, 44, 48, 52, 56, 60, 64, 100, 104, 108, 112, 116, 120, 124, 128, 132, 136, 140, 149, 153, 157, 161, and 165
Maximum Power of Detector Radio 50 mW
Number of Antennas on Detector
2
IP Addressing Static private IP addresses for detectors and access point (AP)
Agency approvals FCC Part 15
Typical Data Size One 15 MB file per image
Dual Homed PC (two NIC cards) Hospital network connection, private network connection
DRX Plus 3543 Detector
1–4 AC3451 | 2016-04-29
Security
WPA2-PSK AES Factory and user loaded keys
SSID Broadcast
Private Patient Identification Data No patient ID data exchanged with detector
User name and Password Non-default user name and password
Enterprise Wireless
IP Addressing Static or DHCP
Authentication The detector supports authentication with most radius servers using the following protocols: EAP-TLS, EAP-PEAP-MSCHAPV2, EAP-FAST, EAP-PEAP-TLS, EAP-TTLS, EAP-TTLS-MSCHAPV2
Encryption Methods AES-CCMP, WPA2-AES
Certificates supported .pern, .crt, .cer, .der, .p7b, .pfx, .p12 extensions
Keys Private keys with .key extensions
QOS Control signals and data path can be separately configured.
AC3451 | 2016-04-29 2–1
2 DRX Plus 3543C Detector
Detector
Imaging Specifications
Detector Type Amorphous Silicon on glass—no tiling
Conversion Screen Deposited CsI (TI)
Pixel Size (μm) Resolution Limit (cyc/mm)
139 x 1393.6
Usable Pixel Area (cm) (inch)(pixels)
35.0 x 42.313.8 x 16.62520 x 3032
MTF (%) RQA-3 Beam RQA-5 Beam
0.05 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical
9982594129211511
9983614432231713
DQE (%) ±2 sigma < 10 % RQA-3 Beam RQA-5 Beam
0.78 μGy
2.5μGy
8.0 μGy
0.78 μGy
2.5 μGy
8.0 μGy
0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical
7063504132231610
7265544537302113
7263534538322315
7064554840322416
7063565044362818
6761534842352818
Energy Range (kVp) 40–150
DRX Plus 3543C Detector
2–2 AC3451 | 2016-04-29
Scan Method Progressive
A/D Conversion (bits) 16
Imaging Specifications
Supported Data Interfaces
Wireless 802.11 A, or N, or G
Tethered 10/100/1000 Ethernet
Environmental
Shock High shock tolerance
Temperature Range
Operating +15 to +30° C (+59 to +86° F)
Ambient storage and shipping -10 to +66° C (+14 to +150° F)
Relative Humidity, Non-Condensing %
Operating 10 to 86
Storage 10 to 86
Ingress Protection IP57
Mechanical
Size (cm) 35 x 43 Cassette (ISO 4090)38.35 x 45.95 x 1.47 cm
Weight 3.17 kg (6.99 lb)
Housing Material Aluminum
Sensor Protection Material Carbon fiber and aluminum plate
Weight Limit Applied to a single 4 cm (1.6 in.) point: 114 kg (250 lb)Distributed evenly over the detector area: 170 kg (375 lb)
Detector Battery
Technology • Lithium-polymer Technology
• “Smart” battery technology prevents overcharge
Size 21 x 15 x 0.67 cm
DRX Plus 3543C Detector
AC3451 | 2016-04-29 2–3
Wireless System
Weight 0.4 kg (12.4 oz.)
Voltage / Energy 14.6 to 14.8V DC, 2.1 to 2.5Ah capacity
Battery "Hot Swap" Capability Yes
Charge Capability • 340 maximum images per charge in Direct Connect Mode.
• 14 days per charge in undisturbed Sleep Mode.
Expected Life 500 charge / discharge cycles results in ~80% full charge energy
Medical Safety IEC 62133 -Safety requirements for secondary cells andbatteries containing alkaline or other non-acid electrolytes
Electromagnetic Compatibility IEC 60601-1-2 Ed. 2.1, 2004-Medical Electrical Equipment Electromagnetic Compatibility Requirements and tests, including CISPR 11:1999 + A2:02 emissions Group 1, Class A)
Detector Battery
Technical Specifications
Network Protocol TCP/IP, IPv4/IPv6
Network Type Isolated private wireless LAN (WLAN) Enterprise Wireless
Wireless Protocol 802.11 A, or N, or G
Antenna:Frequency BandAvailable Channels (fixed at installation)
5 and 2.4 GHz1, 5, 9, 13, 36, 40, 44, 48, 52, 56, 60, 64, 100, 104, 108, 112, 116, 120, 124, 128, 132, 136, 140, 149, 153, 157, 161, and 165
Maximum Power of Detector Radio 50 mW
Number of Antennas on Detector
2
IP Addressing Static private IP addresses for detectors and access point (AP)
Agency approvals FCC Part 15
Typical Data Size One 15 MB file per image
Dual Homed PC (two NIC cards) Hospital network connection, private network connection
DRX Plus 3543C Detector
2–4 AC3451 | 2016-04-29
Security
WPA2-PSK AES Factory and user loaded keys
SSID Broadcast
Private Patient Identification Data No patient ID data exchanged with detector
User name and Password Non-default user name and password
Enterprise Wireless
IP Addressing Static or DHCP
Authentication The detector supports authentication with most radius servers using the following protocols: EAP-TLS, EAP-PEAP-MSCHAPV2, EAP-FAST, EAP-PEAP-TLS, EAP-TTLS, EAP-TTLS-MSCHAPV2
Encryption Methods AES-CCMP, WPA2-AES
Certificates supported .pern, .crt, .cer, .der, .p7b, .pfx, .p12 extensions
Keys Private keys with .key extensions
QOS Control signals and data path can be separately configured.
AC3451 | 2016-04-29 3–1
3 DRX Plus 4343 Detector
Detector
Imaging Specifications
Receptor Type Amorphous Silicon on glass—no tiling
Conversion Screen Attached Gd202 S:Tb
Pixel Size (μm) Resolution Limit (cyc/mm)
139 x 1393.6
Usable Pixel Area (cm) (inch)(pixels)
42.5 x 42.516.7 x 16.73056 x 3056
MTF (%) RQA-3 Beam RQA-5 Beam
0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical
9983614228191310
987954362316107
DQE (%) ±2 sigma < 10 % RQA-3 Beam RQA-5 Beam
IEC 62220-1-1, Edition 1.0, 2015 0.97 μGy
3.1μGy
9.92 μGy
0.97 μGy
3.1 μGy
9.92 μGy
0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical
3532272114953
38343025191385
373430252015106
3430251813953
35312520151074
35302520151174
Energy Range 40–150 kVp
DRX Plus 4343 Detector
3–2 AC3451 | 2016-04-29
Scan Method Progressive
A/D Conversion 16 bits
Imaging Specifications
Supported Data Interfaces
Wireless 802.11 A, or N, or G
Tethered 10/100/1000 Ethernet
Environmental
Shock High shock tolerance
Temperature Range
Operating +15 to +30° C (+59 to +86° F)
Ambient storage and shipping -10 to +66° C (+14 to +150° F)
Relative Humidity, Non-Condensing
Operating 10 to 86 %
Storage 10 to 86 %
Ingress Protection IP57
Mechanical
Size (cm) 43 x 43 Cassette (ISO 4090)45.95 x 45.95 x 1.47 cm
Weight 3.63 kg (8.0 lb)
Housing Material Aluminum
Sensor Protection Material Carbon fiber and aluminum plate
Load Limit Applied to a single 4 cm (1.6 in.) point: 114 kg (250 lb)Distributed evenly over the detector area: 170 kg (375 lb)
Detector Battery
Technology • Lithium-polymer Technology
• “Smart” battery technology prevents overcharge
Size 21.0 x 15.0 x 0.67 cm
DRX Plus 4343 Detector
AC3451 | 2016-04-29 3–3
Wireless System
Weight 0.4 kg (12.4 oz.)
Voltage / Energy 14.6 to 14.8V DC, 2.1 to 2.5Ah capacity
Battery "Hot Swap" Capability Yes
Charge Capability • 340 maximum images per charge in Direct Connect Mode.
• 14 days per charge in undisturbed Sleep Mode.
Expected Life 500 charge / discharge cycles results in ~80% full charge energy
Medical Safety IEC 62133 —Safety requirements for secondary cells andbatter-ies containing alkaline or other non-acid electrolytes
Electromagnetic Compatibility IEC 60601–1–2 Ed. 2.1 —Medical Electrical EquipmentElectro-magnetic Compatibility Requirements and tests,including CISPR 11:1999 + A2:02 emissions Group 1, ClassA)
Detector Battery
Technical Specifications
Network Protocol TCP/IP, IPv4, IPv6
Network Type Isolated private wireless LAN (WLAN)Enterprise Wireless
Wireless Protocol 802.11 A, or N, or G
Antenna:Frequency Band Available Channels (fixed at installation)
5 and 2.4 GHz 1, 5, 9, 13, 36, 40, 44, 48, 52, 56, 60, 64, 100, 104, 108, 112, 116, 120, 124, 128, 132, 136, 140, 149, 153, 157, 161, and 165
Maximum Power of Detector Radio 50 mW
Number of Antennas on Detector
2
IP Addressing Static private IP addresses for detectors and access point (AP)
Agency approvals FCC Part 15
Typical Data Size One 18.5 MB file per image
Dual Homed PC (two NIC cards) Hospital network connection, private network connection
DRX Plus 4343 Detector
3–4 AC3451 | 2016-04-29
Security
WPA2-PSK AES Factory and user loaded keys
SSID Broadcast
Private Patient Identification Data No patient ID data exchanged with detector
User name and Password Non-default user name and password
Enterprise Wireless
IP Addressing Static or DHCP
Authentication The detector supports authentication with most radius servers using the following protocols: EAP-TLS, EAP-PEAP-MSCHAPV2, EAP-FAST, EAP-PEAP-TLS, EAP-TTLS, EAP-TTLS-MSCHAPV2
Encryption Methods AES-CCMP, WPA2-AES
Certificates supported .pern, .crt, .cer, .der, .p7b, .pfx, .p12 extensions
Keys Private keys with .key extensions
QOS Control signals and data path can be separately configured.
AC3451 | 2016-04-29 4–1
4 DRX Plus 4343C Detector
Detector
Imaging Specifications
Receptor Type Amorphous Silicon on glass—no tiling
Conversion Screen Deposited CsI (Tl)
Pixel Size (μm) Resolution Limit (cyc/mm)
139 x 1393.6
Usable Pixel Area (cm) (inch)(pixels)
42.3 x 42.316.6 x 16.63032 x 3032
MTF (%) RQA-3 Beam RQA-5 Beam
0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical
9982594129211511
9983614432231713
DQE (%) ±2 sigma < 10 % RQA-3 Beam RQA-5 Beam
IEC 62220-1-1, Edition 1.0, 2015 0.78 μGy
2.5μGy
8.0 μGy
0.78 μGy
2.5 μGy
8.0 μGy
0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical
7063504132231610
7265544537302113
7263534538322315
7064554840322416
7063565044362818
6761534842352818
Energy Range 40–150 kVp
DRX Plus 4343C Detector
4–2 AC3451 | 2016-04-29
Scan Method Progressive
A/D Conversion 16 bits
Imaging Specifications
Supported Data Interfaces
Wireless 802.11A, or N, or G
Tethered 10/100/1000 Ethernet
Environmental
Shock High shock tolerance
Temperature Range
Operating +15 to +30° C (+59 to +86° F)
Ambient storage and shipping -10 to +66° C (+14 to +150° F)
Relative Humidity, Non-Condensing
Operating 10 to 86 %
Storage 10 to 86 %
Ingress Protection IP57
Mechanical
Size (cm) 43 x 43 Cassette (ISO 4090)45.95 x 45.95 x 1.47 cm
Weight 3.74 kg (8.25 lb)
Housing Material Aluminum
Sensor Protection Material Carbon fiber and aluminum plate
Load Limit Applied to a single 4 cm (1.6 in.) point: 114 kg (250 lb)Distributed evenly over the detector area: 170 kg (375 lb)
Detector Battery
Technology • Lithium-polymer Technology
• “Smart” battery technology prevents overcharge
Size 21.0 x 15.0 x 0.67 cm
DRX Plus 4343C Detector
AC3451 | 2016-04-29 4–3
Wireless System
Weight 0.4 kg (12.4 oz.)
Voltage / Energy 14.6 to 14.8V DC, 2.1 to 2.5Ah capacity
Battery "Hot Swap" Capability Yes
Charge Capability • 340 maximum images per charge in Direct Connect Mode.
• 14 days per charge in undisturbed Sleep Mode.
Expected Life 500 charge / discharge cycles results in ~80% full charge energy
Medical Safety IEC 62133:—Safety requirements for secondary cells and batter-ies containing alkaline or other non-acid electrolytes
Electromagnetic Compatibility IEC 60601-1-2 Ed. 2.1, 2004-Medical Electrical Equipment Elec-tromagnetic Compatibility Requirements and tests, including CISPR 11:1999 + A2:02 emissions Group 1, Class A)
Detector Battery
Technical Specifications
Network Protocol TCP/IP, IPv4, IPv6
Network Type Isolated private wireless LAN (WLAN)Enterprise Wireless
Wireless Protocol 802.11 A, or N, or G
Antenna:Frequency Band Available Channels (fixed at installation)
5 and 2.4 GHz 1, 5, 9, 13, 36, 40, 44, 48, 52, 56, 60, 64, 100, 104, 108, 112, 116, 120, 124, 128, 132, 136, 140, 149, 153, 157, 161, and 165
Maximum Power of Detector Radio 50 mW
Number of Antennas on Detector
2
IP Addressing Static private IP addresses for detectors and access point (AP)
Agency approvals FCC Part 15
Typical Data Size One 18.5 MB file per image
Dual Homed PC (two NIC cards) Hospital network connection, private network connection
DRX Plus 4343C Detector
4–4 AC3451 | 2016-04-29
Security
WPA2-PSK AES Factory and user loaded keys
SSID Broadcast
Private Patient Identification Data No patient ID data exchanged with detector
User name and Password Non-default user name and password
Enterprise Wireless
IP Addressing Static or DHCP
Authentication The detector supports authentication with most radius servers using the following protocols: EAP-TLS, EAP-PEAP-MSCHAPV2, EAP-FAST, EAP-PEAP-TLS, EAP-TTLS, EAP-TTLS-MSCHAPV2
Encryption Methods AES-CCMP, WPA2-AES
Certificates supported .pern, .crt, .cer, .der, .p7b, .pfx, .p12 extensions
Keys Private keys with .key extensions
QOS Control signals and data path can be separately configured.
AC3451 | 2016-04-29 5–1
5 DRX Plus 4343F Detector
Imaging Specifications
Receptor Type Amorphous Silicon on glass—no tiling
Conversion Screen Attached Gd202 S:Tb
Pixel Size (μm) Resolution Limit (cyc/mm)
139 x 1393.6
Usable Pixel Area (cm) (inch)(pixels)
42.5 x 42.516.7 x 16.73056 x 3056
MTF (%) RQA-3 Beam RQA-5 Beam
0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical
9983614228191310
987954362316107
DQE (%) ±2 sigma < 10 % RQA-3 Beam RQA-5 Beam
IEC 62220-1-1, Edition 1.0, 2015 0.97 μGy
3.1μGy
9.92 μGy
0.97 μGy
3.1 μGy
9.92 μGy
0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical
3532272114953
38343025191385
373430252015106
3430251813953
35312520151074
35302520151174
Energy Range 40–150 kVp
DRX Plus 4343F Detector
5–2 AC3451 | 2016-04-29
Scan Method Progressive
A/D Conversion 16 bits
Imaging Specifications
Supported Data Interfaces
Tethered 10/100/1000 Ethernet
Environmental
Shock High shock tolerance
Temperature Range
Operating +15 to +30° C (+59 to +86° F)
Ambient storage and shipping -10 to +66° C (+14 to +150° F)
Relative Humidity, Non-Condensing
Operating 10 to 86 %
Storage 10 to 86 %
Ingress Protection IPX0
Mechanical
Size (cm) 43 x 43 Cassette (ISO 4090)45.95 x 45.95 x 1.47 cm
Weight 3.53 kg (7.8 lb)
Housing Material Aluminum
Sensor Protection Material Carbon fiber and aluminum plate
Load Limit Applied to a single 4 cm (1.6 in.) point: 114 kg (250 lb)Distributed evenly over the detector area: 170 kg (375 lb)
Security
Private Patient Identification Data No patient ID data exchanged with detector
User name and Password Non-default user name and password
AC3451 | 2016-04-29 6–1
6 DRX Plus 4343FC Detector
Imaging Specifications
Receptor Type Amorphous Silicon on glass—no tiling
Conversion Screen Deposited CsI (Tl)
Pixel Size (μm) Resolution Limit (cyc/mm)
139 x 1393.6
Usable Pixel Area (cm) (inch)(pixels)
42.3 x 42.316.6 x 16.63032 x 3032
MTF (%) RQA-3 Beam RQA-5 Beam
0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical
9982594129211511
9983614432231713
DQE (%) ±2 sigma < 10 % RQA-3 Beam RQA-5 Beam
IEC 62220-1-1, Edition 1.0, 2015 0.78 μGy
2.5μGy
8.0 μGy
0.78 μGy
2.5 μGy
8.0 μGy
0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical
7063504132231610
7265544537302113
7263534538322315
7064554840322416
7063565044362818
6761534842352818
Energy Range 40–150 kVp
DRX Plus 4343FC Detector
6–2 AC3451 | 2016-04-29
Scan Method Progressive
A/D Conversion 16 bits
Imaging Specifications
Supported Data Interfaces
Tethered 10/100/1000 Ethernet
Environmental
Shock High shock tolerance
Temperature Range
Operating +15 to +30° C (+59 to +86° F)
Ambient storage and shipping -10 to +66° C (+14 to +150° F)
Relative Humidity, Non-Condensing
Operating 10 to 86 %
Storage 10 to 86 %
Ingress Protection IPX0
Mechanical
Size (cm) 43 x 43 Cassette (ISO 4090)45.95 x 45.95 x 1.47 cm
Weight 3.63 kg (8.0 lb)
Housing Material Aluminum
Sensor Protection Material Carbon fiber and aluminum plate
Load Limit Applied to a single 4 cm (1.6 in.) point: 114 kg (250 lb)Distributed evenly over the detector area: 170 kg (375 lb)
Security
Private Patient Identification Data No patient ID data exchanged with detector
User name and Password Non-default user name and password
AC3451 | 2016-04-29 7–1
7 DRX Plus Statements Related to Systems Integration
Detector Safety
The detector has been approved to IEC 60601-1 by TUV Rheinland and has a cTUVus and CE safety certification as a Medical Electrical Equipment device.
Recycling
The detector should be returned to Carestream Health at the end of life. The components will be recycled appropriately.
The Lithium-polymer battery is WEEE compliant (as shown by the Wheelie Dustbin logo) and should be recycled at local recycling stations. Carestream Health pays taxes in the EU Battery Directive countries to cover the recycling costs.
Line Noise Correction
The Image Correction Preference is used for the following purpose:
• Normal—Reduces correlated line noise due to readout ICs. This is a very subtle correction, limited to near zero signal levels.
• Enhanced—Reduces correlated line noise due to external EMI sources, such as motor drives and power supplies. This level should only be used if there is detectable line noise in the images.
Grids
• Moving Grids: Moving grids should not exhibit grid aliasing. If the exposure time is very short, the grid may alias with any digital detector. This is normally not a problem, since grids are used on large body parts, which generally require exposure times longer than 20 ms.
• Low Resolution Stationary Grids: 30-to-50 Lines/cm (76-to-127 Lines/inch) Carestream image processing software will automatically remove the visible lines from the full resolution image.
• High Resolution Stationary Grids: ≥ 75 Lines/cm (>191 Lines/inch)—Grid resolutions of 75 Lines/cm or greater will not be visible on images.
• Medium Resolution Stationary Grids: 50-to-75 Lines/cm (127-to-191 Lines/inch)—Stationary grids in this range may cause aliasing and are therefore not recommended. However, medium resolution moving grids should not exhibit aliasing.
DRX Plus Statements Related to Systems Integration
7–2 AC3451 | 2016-04-29
Durability
The detector has been designed to be a robust device and function properly in normal handling. The instrument may be damaged if it receives a strong jolt from a drop or strike. The warranty does not cover damage as a result of accident or abuse, including broken detector glass and other obvious damage, however a drop rider can be purchased to protect the investment. The detector has been equipped with internal accelerometers that will register when the detector has incurred a significant shock. The drop height that results in a registered event depends on the orientation of the detector during the drop, and the stiffness of the surface it strikes.
Detector Life
The detector system is specified to last 10 years under normal use of 56,000 images per year. That is ~560K images. This statement relates to the life of the physical aspects of the product, including hardware, normal physical handling and image acquisition
No Preventive Maintenance (PM) is required on the system.
Detector Quality Assurance
The detector must undergo gain calibration on an annual basis. A 360 day reminder is built into the software.
Express Calibration restores Gain Calibration and Defect Calibration and does a defect quality check. Full Calibration is the same as Express Calibration and adds quality checks for Signal Uniformity, Noise Uniformity and Dark Noise
Permanent pixel defects are identified in the factory based on outlier performance in dark noise, X-ray uniformity response, and time lag response. Additional pixel defects are identified during Express Calibration and Full Calibration
Quality Limits
• No 4 x 4 defect clusters.
• Signal Uniformity = 2 %
• Noise Uniformity = 40 % (lowest exposure), 31% (1/3 gain calibration exposure), 28 % (gain calibration exposure), 28 % (2 times gain calibration exposure)
• Two Dark Noise = 9.2 ADC
Wireless FAQs
What impact will the system AP have on the Hospital/Clinic WLAN?
The system AP will have the channel fixed and the transmit power reduced to the minimum level needed for reliable operation of the detector moving around the room in which the AP is installed. Because of the intermittent bursts of data lasting only a few seconds, and the low transmission power, the impact of the system WLAN on the Hospital mobile devices is minimized.
DRX Plus Statements Related to Systems Integration
AC3451 | 2016-04-29 7–3
What is the range of the wireless communication?
The wireless communication should have plenty of range for any single room installation. The connection range is typically more than 30 m, depending on the obstructions between the access point and the detector
DRX Plus Statements Related to Systems Integration
7–4 AC3451 | 2016-04-29
AC3451 | 2016-04-29 I
Publication History
Version Date Changes
A 2015-06-16 First release
B 2016-02-10 Updated the specifications for MTF and DQE.
C 2016-04-29 Added information for 4343 detectors.