Principles of Imaging Science II (RAD 120)
Radiographic Grids
1
Image-Forming X-Rays
Four X-ray paths • a. X-rays interact with patient and
scatter away from the receptor
• b. X-rays interact and are absorbed (photoelectric absorption) within patient
• c. X-rays are transmitted through patient without interaction and strike receptor
• d. X-rays interact with patient (Compton scatter) and scatter towards
• C and D are referred to the image- forming x-ray photons
2
Radiographic Grid
• Used to reduce scatter radiation from reaching the image receptor (IR) through absorption
• Cleans up scatter radiation
• Inherent part of bucky, placed between the patient and IR
• Table or upright bucky usage
– >60 kVp, 10 cm tissue
• When primary x-rays interact
with the patient, x-rays are scattered from the patient in all directions.
3
2
Image Contrast • Difference in optical
density between adjacent structures
• High vs Low contrast
– Skeletal anatomy
– Abdomen, Chest
High
Medium
Low
Radiographs of a cross section of long bone. A, High
contrast would result from the use of only transmitted,
unattenuated x-rays. B, No contrast would result from the
use of only scattered x-rays. C, Moderate contrast results
from the use of both transmitted and scattered x-rays. 4
Grid Design
• Radiolucent interspace material with alternating radiopaque strips – Aluminum, plastic or carbon fiber
for interspace
– Lead, tungsten, platinum, gold strips
• Transmits x-rays traveling in a straight line, oblique x-rays absorbed by strips
The only x-rays transmitted through a grid
are those that travel in the direction of the
interspace. X-rays scattered obliquely
through the interspace are absorbed.
5
% Grid Surface X-ray Absorption
• Formula applied to determine the percentage of x-rays exiting the body that will be absorbed
• Based upon grid design – Lead strip width and interspace width
– Higher % yields > absorption
Surface area
of grid 6
3
Application
7
Grid Ratio
• Grid Ratio is the height of the grid strip (h) divided by the thickness of the interspace material (D). T = strip width. – Grid Ratio = h/D
• Affected by changing – Height of lead strips
– Thickness of strips
– Width of interspace
8
Grid Ratio
• High ratio grids absorb more scatter yet require higher mAs or kVp
– mAs is factor of choice
• 5:1, 6:1, 8:1, 10:1, 12:1, 16:1 ratio designs
High-ratio grids are more effective than
low-ratio grids because the angle of
deviation is smaller.
9
4
Grid Ratio Application
10
Grid Frequency
• Number of strips or grid lines per inch or cm – 25 – 45 lines/cm, 60 – 110
lines/in – 25 – 80 lines/cm, 60 – 200
lines/in
• Higher grid frequency requires higher technique – Less grid lines appear in image – Often used in mammography
• 80 lines/cm, 200 lines/in
• Typically higher frequency grids have thinner lead strips
11
Digital Imaging Systems
• Very high-frequency grids
– 103-200 lines/in
– 41-80 lines/cm
• Recommended for use with digital systems
– Minimizes grid line appearance
12
5
Lead Content of Grid
• Lead content
– Most important factor in grid’s efficiency
– Measured in mass per unit area
• g/cm2
– High ratio grids tend to have highest lead content
– As lead content increases, removal of scatter increases and therefore contrast increases
13
Application
14
Grid Performance
• Contrast Improvement Factor (k)
– Comparison of image contrast with a grid to image contrast without a grid
– k is higher for higher ratio grids
• K = Radiographic contrast with grid
Radiographic contrast without grid
• Measured at 100 kVp using a step wedge
• Manufacturer Avg 1.5 – 2.5 – Use of a grid approximately doubles the contrast
15
6
Grid Performance • Bucky Factor
– Higher technique required with grid usage
• mAs X bucky factor avg
– Measurement of technical factor and patient dose increase based upon penetration of primary & scatter radiation through the grid
– Bucky factor increases with increased grid ratio and increased kVp
Grid Ratio
70 kVp
90 kVp
120 kVp
Avg
Non-Grid
1 1 1 1
5:1 2 2.5 3 2
8:1 3 3.5 4 4
12:1 3.5 4 5 5
16:1 4 5 6 6
Approximate Bucky Factor Values
16
Approximate Exposure Factor Changes Necessary
for Standard Grids
Grid Ratio mAs Increase kVp Increase
Non-Grid 1X 0
5:1 2X 8 – 10
6:1 3X 11 – 12
8:1/10:1 4X 13 – 15
12:1 5X 20 – 25
16:1 6X 30 – 40
17
Grid Types Linear Parallel Linear Focused
• Linear/Parallel – Vertical lead strips do not
coincide with the primary beam
– Absorption of 10 beam (Grid cut-off) occurs with:
• Short SID
• Large IR
A parallel grid is constructed with parallel
grid strips. At a short source-to-image
receptor distance (SID), some grid cutoff
may occur.
18
7
Parallel Grid
With a parallel grid, optical density
(OD) decreases toward the edge of the
image receptor. The distance to grid
cutoff is the source-to-image receptor
distance (SID) divided by the grid ratio. 19
Grid Types
• Linear/Focused
– Angled lead strips to coincide with primary beam divergence
– Focal distance set to SID usage to minimize grid cut-off
20
Comparison of Transmitted Photons Parallel & Focused Grids
21
8
Grid Focus
Convergent Line.
Imaginary lines drawn above a
linear focused grid from each
lead strip meet to form a
convergent point. The points
form a convergent line along the
length of the grid.
Convergent Point.
The convergent line or point of a
focused grid falls within a focal
range.
Grid Types
• Crossed (Criss-Cross, Cross-Hatched)
– 2 parallel grids perpendicular
– Not common
– High Grid cut-off if off-centered to CR
23
Grid Types
• Moving grid mechanism – Upright or table bucky activation
– Eliminate grid lines from image
• High frequency = less grid lines possible
• Low frequency = more grid lines possible
– Reciprocating
• 2 cm movement transversely
• Motor drives grid back and forth during exposure
– Oscillating
• Circular movement 2-3 cm movement
• Electromagnet pulls grid to one side
• Releases it during exposure
24
9
Grid Types
• Grid Cap
– Permanently mounted grid, IR slides into device
– Used with a variety of IR sizes
• Grid Cassette
– Permanently mounted grid
– Specific grid sizes
• Wafer Grid
– Non-permanent grid mount, must be secured
– Specific grid size
25
Long vs. Short Dimension Grids
• Orientation of lead strips for a long- and short-dimension grid.
Grid Types
• Air-Gap (Air Filtration) – Common on dedicated
Chest X-ray units
– Part is @10-15 cm from IR ( 4” - 6”)
– Similar to 8:1ratio grid
• 10” air gap equivalent to 15:1 ratio grid
– mAs increased 10% per cm gap
– Magnification results unless SID is increased
6”
27
10
Grid Cutoff
• A decrease in the number of transmitted photons that reach the image receptor because of some misalignment of the grid
• Grid Errors
– Off-level grid
– Off-center grid
– Off-focus grid
– Upside-down focused grid
Grid Errors • Off-level
– Parallel & focused
– Decreased density across image
29
Grid Errors
• Off-Center (lateral decentering)
– Focused Grid
– Decreased density across image
– Most common error
30
11
Grid Errors • Off-focus
– Incorrect SID use
– Decreased density at edges of image • Direct relationship
– More critical with high ratio grid
• Upside-down
– Mobile radiography
– CR not directed to tube side
– Marked decreased density at edges of image and points lateral to CR
31
Grid Cutoff – Off Level
32
Grid Cutoff – Off Center
33
12
Grid Cutoff – Off Focus
34
Grid Cutoff – Upside Down Focused
35
Grid Cutoff
Upside-Down Focused Grid Cutoff.
Radiograph produced with an
upside-down focused grid
Off-Center Grid Cutoff.
Radiograph demonstrating grid cutoff
caused by off-centering.
13
Grid Errors - Moire Effect • Zebra pattern
• Caused by
– Similar grid frequency to laser scanning frequency in CR processing
– Using a grid cassette in the bucky tray
• Correct by
– Selecting a high grid frequency
– Use a moving grid mechanism
– Do not use two grids
37
Grid Selection • Patient dose
• Kvp usage
• Scatter absorption
– <90 kVp 8:1 satisfactory
– >90 kvp >8:1 grids used
As grid ratio increases, transmission of scatter radiation decreases
faster than transmission of primary radiation. Therefore, cleanup of
scatter radiation increases. 38
Clinical Consideration in Grid Selection
Grid Degree of Scatter Removal
Off-Center latitude
Off-focus latitude
kVp
Comments
5:1 + Very Wide Very Wide Up to 80 Low cost; easy to use
6:1 + Very Wide Very Wide Up to 80 Low cost; mobile radiography
8:1/10:1 +/+++ Wide/Wide Wide/Wide Up to 100 General stationary exams
12:1 ++++ Narrow Narrow Over 110 Precise centering; usually fixed mount
16:1 +++++ Narrow Narrow Over 100 Precise centering; usually fixed mount