Pulse Sequences 120505095204 Phpapp02

Post on 05-Mar-2016

220 views 0 download

description

Pulse Sequences 120505095204 Phpap

Transcript of Pulse Sequences 120505095204 Phpapp02

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 1/82

MRI Pulsesequences

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 2/82

1. SPIN ECHO (SE )

2. FAST SPIN ECHO (FSE)

3. INVERSION RECOVERY ( STIR, FLAIR)

4. GRADIENT READING ECH0 ( GRE )

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 3/82

The spin echo sequence

The spin echo sequence is a very basic and sequence

The two main parameters of a spin echo (SEare the TR and the TE.

TR is the time between two 90° excitation puregular SE sequences, TR is about 100 to 30

TR is the time allowed for longitudinal magnerecover (T1 relaxation). The longer the TR iscomplete the longitudinal magnetization recoIf the TR is short, not all tissues will have und

complete T1 relaxation and image contrast wmore dependent on the T1 relaxation proces

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 4/82

With a regular SE sequence, only one line of k-sduring a repetition (which lasts for time TR). Theacquisition (TA) of the full k-space is therefore pthe TR (TA = TR * number of lines).

TE is the time between the 90° excitation pulse signal acquisition. The echo concerns transversmagnetization and is the result of spin rephasin180° RF pulse applied at time TE/2. The 180° Rresponsible for a cancelation of spin dephasingfield inhomogeneities so that the transverse madecay is T2-dependent instead of T2*-dependen

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 5/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 6/82

Spin Echo

SE sequences serve as reference for tissue con

Pros:

 Anatomic imaging

Reference for tissue signal and image contrast

Cons:

Very long acquisition time

Clinical use:  Almost all the organs explored in MRI

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 7/82

T1-weighted

If the TR is very long, the longitudinal magnetization tissues will have recovered completely (complete T1

On the contrary, if the TR is short, tissue signal and

will depend on the T1 characteristics of tissues as nowill have completely recovered their longitudinal magIf the TE is also short, there is little time for T2 relaxacontrast will not be very dependent on T2.

When both the TR and the TE are short, the image isweighted.

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 8/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 9/82

Spin Echo

T1-weighted

Pros:

 Anatomic imaging

Cons:

Very long acquisition time

Clinical use:

 Almost all the organs explored in MRI

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 10/82

T2-weighted

To obtain a T2-weighted image in SE, tmust be long so that the longitudinalmagnetization of the tissues has time tocompletely and the TE must be long to

relaxation happen.

 A long TR is about 2000 ms or more anTE is about 80 to 140 ms.

With such parameters, the tissues withtime will have a stronger signal than thewith a short T2 time.

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 11/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 12/82

T2-weighted

Pros:

Imaging of water/fluids (CSF, edema, b

Cons:

Very long acquisition time

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 13/82

PD-weighted

If we use a long TR (> 2000 ms) and a (10 - 20 ms), the tissue signal will not bdependent on T1 and T2 relaxation.

Therefore, image contrast is mainly duedensity.

Tissues with a rich Hydrogen content (will be bright whereas tissues with low

content will be dark.

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 14/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 15/82

PD-weighted

Cons:

Very long acquisition time

Clinical use:

Osteoarticular 

Pediatric Neuroradiology

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 16/82

2.Fast Spin Echo ( Turbo Spin  A regular SE sequence requires as many repetitions

lines in the k-space to complete a slice acquisition.

Instead of acquiring k-space lines of other slices at d

positions during the wasted time, it is possible to acqspace lines for the same slice.

Such fast spin echo sequences use one 90° excitatiotwo or more 180° pulses in the same repetition time different phase-encoding gradient steps, to acquire m

that will fill the k-space.

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 17/82

The number of echoes acquired after a singexcitation is called the Turbo Factor or EchoLength.

 As each echo undergoes more and more T2image contrast is modified.

The effective echo time, determined by whelines of k-space are acquired, indicates appwhat the image contrast obtained is like (ratweighted or T2-weighted).

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 18/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 19/82

Pros:

Reduction of scan time

Low sensitivity to magnetic susceptibili

Cons:

Modification of tissue contrast

Clinical use:

 Almost all the organs explored in MRI

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 20/82

Single Shot Spin Echo

Single Shot Spin Echo, Rare

If there are as many echoes as the numb

space lines to fill, the entire k-space can

acquired after a unique 90° excitation puis called a single-shot sequence.

The resulting images are strongly T2-we

most of the k-space lines are acquired w

TE.

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 21/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 22/82

Rare Pros:

Excellent contrast between tissues and still fluids

Very fast : useful for moving organs

Cons:

Decreased signal to noise ratio

Low spatial resolution

Clinical use:

Biliary and urinary tract exploration ,myelograph

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 23/82

HASTE, SSFSE, SSTSE

Single Shot Spin Echo with Partial Filling of K-Space

Commercial name: HASTE (Siemens), SSFSE (GE), SST

Pros:Excellent contrast between tissues (gray or dark) and still fl

Very fast : less sensitive to motion, breath-hold sequences

Cons:

Decrease of signal to noise ratio (low signal amplitude of laeffective TE, partial filling of k-space)

Low spatial resolution

Clinical use: Hepato-biliary and urinary tract exploration

Cardiac MRI without gating

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 24/82

Inversion Recovery

One method for manipulating contrast is called Inversion-R

 An Inversion-Recovery consists in applying a 180° inversiobeginning of the sequence.

The 180° inversion pulse changes the direction of the longmagnetization vector to its opposite. Then, the longitudinagoing to recover as defined by T1 relaxation.

 At time TI (inversion time), a regular spin echo (or gradienecho planar) sequence is performed, starting with an excit

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 25/82

The TI is defined so that the longitudinal magnetizationtissue is null.

Consequently, this tissue will have a null transverse mafter the excitation pulse, resulting in a signal suppresstissue.

The optimal TI for eliminating the signal of a given tisson the tissue's characteristic T1 time.

The displayed image usually shows the magnitude of which corresponds to the absolute value of the signal isigned value

Tissues with no magnetization appear in dark and tissumagnetization (positive or negative) appear in gray or

Inversion-Recovery allows to eliminate the signal of tisaccording to their T1 time by choosing an appropriate T

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 26/82

Inversion Recovery

Fat Signal Suppression STIR

Fat : T1 time is short, And hence fat signal can be elimin

an inversion-recovery with a short

(about 140 milliseconds).

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 27/82

STIR  Pros:

Fat signal suppression

Cons:

Increased scan time

Clinical use:

MSK SYSTEM

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 28/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 29/82

Water Signal Suppression

FLAIR  In the same way, with a TI time ab

ms, water signal is eliminated (wa

long).

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 30/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 31/82

FLAIR  Pros:

Elimination of CSF signal

Useful with T2-weighted images (edem

Cons:

Long scan time

Clinical use:

Neuroradiology

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 32/82

4. Gradient Echo

Two major differences distinguish the gradietechnique from the spin echo technique:

 An excitation pulse with a flip angle lower th

No 180° rephasing pulse

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 33/82

 A flip angle lower than 90° (partial flip anglethe amount of magnetization tipped into the plane.

The consequence of a low-flip angle excitat

faster recovery of longitudinal magnetizationshorter TR/TE and decreases scan time.

The advantages of low-flip angle excitationstechniques are faster acquisitions, new contbetween tissues and a stronger MR signal inshort TR.

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 34/82

In the following eg., there differences in longitudinal and trans

magnetization after a 90° excitation pulse and a 30° excitatio

partial flip angle excitation, the decrease in longitudinal magnvery low (only a 13 percent loss) but the transverse magnetiz

of its maximum value (sin 30°).

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 35/82

Gradient Echo

The flip angle determines the fraction of magnetin the transverse plane (which will produce the Nand the quantity of magnetization left on the lon

If the flip angle decreases, the residual longitud

magnetization will be higher and the recovery ofmagnetization for a given T1 and TR will be mo

On the other hand, the result of a lower flip angis a lower tipped magnetization.

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 36/82

Gradient Echo

The actual decay of the transverse magnetizatioseveral factors:

spin-spin tissue-specific relaxation (T2) which iB0 field inhomogeneities and magnetic suscwhich are static

 As GE techniques use a single RF pulse and norephasing pulse, the relaxation due to fixed caureversed and the loss of signal results from T2* T2 + static field inhomogeneities).

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 37/82

Gradient Echo

Signal weighting in GE imaging lies on 3 parameters

TR

TE

Flip angle

The resulting contrast in basic GE sequences is a vaT1 and T2*:

the higher the flip angle chosen, the more T1-weightwill be

the shorter the TE obtained, the less T2*-weighted th

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 38/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 39/82

Gradient Echo

There is an optimal combination between th

the flip angle so that the NMR signal is max

The optimal flip angle is called the Ernst ang

It is calculated from the TR value and the T1

specific value to give the best flip angle to c

want to obtain the maximal signal for a give

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 40/82

Gradient Echo

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 41/82

If you compare it to an SE sequence dican note that a GE sequence is simplesingle RF pulse, which allows for shorteTE and faster acquisition.

Data is acquired during the gradient ecby the bipolar readout gradient (with a lobe).Note this is the same principle with datacquisition in SE sequences (but the spand the gradient reading echo match)

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 42/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 43/82

Gradient Echo

Pros:fast technique

Cons: T2*-weighted images instead of T2

More sensitive to magnetic susceptibili

Clinical use:

eg. Hemorrhage , calcification

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 44/82

5. Advanced Gradient Echo

Steady-state

The GE technique allows for very short TR and TE.

The TR can be so short that the spins on the slice plhave enough time to dephase completely the MR sigdecays completely. These sequences are called stea

They require that TR is less than T2*.

The steady-state technique produces T2*-weighted ifast (in less than 1 second with TRs < 10 msec).

The contrast between tissues and fluids is very high have a good signal to noise ratio.

 As TEs are also very short, blood generally appears

does not have enough time to move out of the slice p

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 45/82

Advanced Gradient Echo

Steady-state In order to maintain the steady-state,

shifts from one excitation to the next mremain constant.

This is done by applying a phase-encrewinder gradient after the signal read

The rewinder gradient is of equal ampopposite sign of the phase-encoding g

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 46/82

It restores the original phase state before phencoding, which is needed for spatial localizsequences are called steady-state GRE.

Image contrast resulting from steady-state ismix of T1 and T2* so that steady-state can bwe want to obtain T1-weighted images.

In these cases, a spoiler gradient or RF is udestroy the transverse steady-state. These are called spoiled GRE sequences.

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 47/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 48/82

Advanced Gradient EchoSteady-state

Commercial name: FISP (Siemens), FFE (Philips),

Pros:

 Accentuated signal form liquids

Very good contrast between liquids and soft tissues

Very fast (breath-hold sequences, dynamic cine imag

Cons:

Image contrast and sensitivity to motion very compleon TR, TE and flip angle

Clinical use:

urinary tract and biliary explorations

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 49/82

Advanced Gradient Echo

Enhanced Steady-state

Commercial name: True-FISP (Siemens), Balance(Philips), FIESTA (GE).

Using a fully-balanced gradient waveform, the statio

returns to the same phase it had before the gradienapplied.

This pulse sequence produces images with increasfluid.

True FISP is a reliable technique, very fast and relainsensitive, with very good contrast between liquidstissues.

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 50/82

Pros:

Increased signal from fluid

Very good contrast between liquids and

Very fast (motion imaging)

Clinical use:

urinary tract, pelvic and biliary exploratio

Pediatrics and antenatal imaging ++

Cardiac and vascular imaging ++

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 51/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 52/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 53/82

Advanced Gradient EchoSpoiled gradient echo

Commercial name :SPGR (GE), FLASH (Siemens), FFE

Pros:

T1 weighted anatomic imaging

Short scan time, volume acquisitions, breath-hold sequenc

MR angiography

Cons:

Magnetic susceptibility artifacts

Clinical use:

Cardio-vascular MRI

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 54/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 55/82

Image quality and artifact

The quality of an MR image depends on seve

1. Spatial resolution and image contrast

2. Signal to noise ratio (and contrast to noise rat3.  Artifacts

 An MR exploration is a compromise between image quality.

 An MR exploration protocol and its sequence will have to be optimized in function of the orgpathology.

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 56/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 57/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 58/82

Spatial resolution Spatial resolution corresponds to t

of the smallest detectable detail.

The smaller the voxels are, the hig

potential spatial resolution will be.

Voxel volume is determined by the

size (256 x 256 or 512 x 512 etc..)

of view (10 cm, 20 cm, etc....), andthickness.

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 59/82

FOV 25

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 60/82

FOV 35

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 61/82

FOV 50

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 62/82

MATRIX SIZE (128 X 128

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 63/82

MATRIX SIZE (256 X 256

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 64/82

MATRIX SIZE (512 X512)

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 65/82

Signal and Noise

Noise is like interferences which present as a irregranular pattern.

This random variation in signal intensity degradeinformation.

The main source of noise in the image is the pati

(RF emission due to thermal motion). The whole measurement chain of the MR scanne

electronics...) also contributes to the noise.

This noise corrupts the signal coming from the tramagnetization variations of the intentionally excitthe selected slice plane).

The signal to noise ratio (SNR) is equal to the rataverage signal intensity over the standard deviatnoise.

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 66/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 67/82

INCREASING NOISE

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 68/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 69/82

Signal and Noise

The signal to noise ratio depends both on someare beyond the operator's control (the MR scanspecifications and pulse sequence design) and that the user can change:

Fixed factors : static field intensity, pulse seqtissue characteristics

Factors under the operator's control

RF coil to be used

Sequence parameters :

voxel size (limiting spatial resolution), numbeaveragings, receiver bandwidth

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 70/82

Surface coil The smaller the sensitive volume

the lower the noise from the adjac

structures of the selected slice pla

it can detect, and the better the signoise ratio will be.

 A local coil, or better, a surface co

higher signal to noise ratio than a

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 71/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 72/82

Signal and Noise

The signal comes from the excited protons on thslice plane. The number of spins in parallel stateproportional to the static magnetic field intensity

the field intensity is, the higher the excess numbparallel state (available to make the MR signal) the signal intensity varies almost linearly with thintensity.

 Assuming a uniform proton density, the numberspins is proportional to the voxel size and so is intensity. The signal goes up linearly with the vo

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 73/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 74/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 75/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 76/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 77/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 78/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 79/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 80/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 81/82

7/21/2019 Pulse Sequences 120505095204 Phpapp02

http://slidepdf.com/reader/full/pulse-sequences-120505095204-phpapp02 82/82