Molecular Spectroscopy Visible and Ultraviolet Spectroscopy - UV/VIS Spectroscopy - UV/VIS...

Post on 13-Dec-2015

267 views 7 download

Tags:

Transcript of Molecular Spectroscopy Visible and Ultraviolet Spectroscopy - UV/VIS Spectroscopy - UV/VIS...

Molecular SpectroscopyVisible and Ultraviolet Spectroscopy

- UV/VIS Spectroscopy- UV/VIS Spectrometer- Application for Quantitative Analysis

Ultraviolet: 190~400nm Violet:   400 - 420 nm Indigo:   420 - 440 nm Blue:   440 - 490 nm Green:   490 - 570 nm Yellow:   570 - 585 nm Orange:   585 - 620 nm Red:   620 - 780 nm

Internal Energy of Molecules

Etotal=Etrans+Eelec+Evib+Erot+Enucl

Eelec: electronic transitions (UV, X-ray)

Evib: vibrational transitions (Infrared)

Erot: rotational transitions (Microwave)

Enucl: nucleus spin (nuclear magnetic

resonance) or (MRI: magnetic resonance

imaging)

Electronic Spectroscopy

Ultraviolet (UV) and visible (VIS) spectroscopy This is the earliest method of molecular

spectroscopy. A phenomenon of interaction of molecules with

ultraviolet and visible lights. Absorption of photon results in electronic

transition of a molecule, and electrons are promoted from ground state to higher electronic states.

UV and Visible Spectroscopy

In structure determination : UV-VIS spectroscopy is used to detect the presence of chromophores like dienes, aromatics, polyenes, and conjugated ketones, etc.

Electronic transitions

There are three types of electronic transition

which can be considered; Transitions involving , , and n

electrons Transitions involving charge-transfer

electrons Transitions involving d and f electrons

Absorbing species containing , , and n electrons

Absorption of ultraviolet and visible radiation in organic molecules is restricted to certain functional groups (chromophores) that contain valence electrons of low excitation energy.

UV/VIS

Vacuum UV or Far UV (λ<190 nm )

Transitions

An electron in a bonding s orbital is excited to the corresponding antibonding orbital. The energy required is large. For example, methane (which has only C-H bonds, and can only undergo transitions) shows an absorbance maximum at 125 nm. Absorption maxima due to transitions are not seen in typical UV-VIS spectra (200 - 700 nm)

n Transitions

Saturated compounds containing atoms with lone pairs (non-bonding electrons) are capable of n transitions. These transitions usually need less energy than transitions. They can be initiated by light whose wavelength is in the range 150 - 250 nm. The number of organic functional groups with n peaks in the UV region is small.

n and Transitions

Most absorption spectroscopy of organic compounds is based on transitions of n or electrons to the excited state.

These transitions fall in an experimentally convenient region of the spectrum (200 - 700 nm). These transitions need an unsaturated group in the molecule to provide the electrons.

Chromophore Excitation max, nm Solvent

C=C →* 171 hexane

C=On→*→*

290180

hexanehexane

N=On→*→*

275200

ethanolethanol

C-X   X=Br, I

n→*n→*

205255

hexanehexane

Terms describing UV absorptions

1. Chromophores: functional groups that give

electronic transitions.

2. Bathochromic shift: shift to longer λ, also called red shift.

3. Hysochromic shift: shift to shorter λ, also called blue shift.

4. Hyperchromism: increase in ε of a band.

5. Hypochromism: decrease in ε of a band.

Selection RuleOrbital Spin States & Multiplicity

Singlet state (S):Most molecules have ground state with all electron spin paired and most excited state also have electron spin all paired, even though they may be one electron each lying in two different orbital. Such states have zero total spin and spin multiplicities of 1, are called singlet (S) states.

Total Spin Multiplicities

Multiplicity Triplet State

For some of the excited states, there are states with a pair of electrons having their spins parallel (in two orbitals), leading to total spin of 1 and multiplicities of 3.

Total Spin Multiplicities

Selection Rule

For triplet state: Under the influence of external field, there are three values (i.e. 3 energy states) of +1, 0, -1 times the angular momentum. Such states are called triplet states (T).

According to the selection rule, S→S, T→T, are allowed transitions, but S→T, T→S, are forbidden transitions.

Selection Rules of electronic transition

Electronic transitions may be classed as intense or weak according to the magnitude of εmax th

at corresponds to allowed or forbidden transition as governed by the following selection rules of electronic transition:

Spin selection rule: there should be no change in spin orientation or no spin inversion during these transitions. Thus, S→S, T→T, are allowed, but S→T, T→S, are forbidden. ( S=0 △ transition allowed)

Instrumentation

光源 分光器 樣品 偵測器 記錄器

Components of a SpectrophotometerLight Source

Deuterium Lamps- a truly continuous spectrum in the ultraviolet region is produced by electrical excitation of deuterium at low pressure. (160nm~375nm)

Tungsten Filament Lamps- the most common source of visible and near infrared radiation.

Components of a Spectrophotometer

Monochromator (分光器 /單光器 )

Used as a filter: the monochromator will select a narrow portion of the spectrum (the bandpass) of a given source

Used in analysis: the monochromator will sequentially select for the detector to record the different components (spectrum) of any source or sample emitting light.

MonochromatorCzerny-Turner design

Grating

Photomultiplier Detector

Principle of Photomultiplier Detector

The type is commonly used. The detector consists of a photoemissive catho

de coupled with a series of electron-multiplying dynode stages, and usually called a photomultiplier.

The primary electrons ejected from the photo-cathode are accelerated by an electric field so as to strike a small area on the first dynode.

Principle of Photomultiplier Detector

The impinging electrons strike with enough energy to eject two to five secondary electrons, which are accelerated to the second dynode to eject still more electrons.

A photomultiplier may have 9 to 16 stages, and overall gain of 106~109 electrons per incident photon.

Single and Double Beam Spectrometer

Single-Beam: There is only one light beam or optical path from the source through to the detector.

Double-Beam: The light from the source, after passing through the monochromator, is split into two separate beams-one for the sample and the other for the reference.

Quantitative AnalysisBeer’s Law

A=bc

: the molar absorptivity (L mol-1 cm-1)

b: the path length of the sample

c :the concentration of the compound in solution, expressed in mol L-1

Transmittance

I0 I

b

303.2

log)log(

)log(303.2)ln(

0

00

00

00

0

k

bcATI

I

I

Ikbc

I

I

dbkcI

dI

kcdbI

dI

I

IT

I

I

b

External Standard and the Calibration Curve

Fluorescence Spectroscopy

Excitation of Molecules

Excitation of molecules can be brought about by absorption of two bands of radiation, one centered about the wavelength 1(S0→S1) and the second around the shorter wavelength 2(S0→S2).

Vibrational Relaxation (VR)

The molecule can rapidly dissipate excess vibrational energy as heat by collision with solvent molecules.

Average lifetime of vibrational relaxation excited molecule is 10-12sec or less.

Fluorescence always involves a transition from the lowest vibrational level of an excited state to any one of the vibrational level of the ground state.

Internal Conversion (IC)

A molecule passes to a lower energy electronic state.

The molecule can pass from a low vibrational level of S2 to an equally energetic vibrational level of the first excited singlet S1.

The mechanism of internal conversion process S2→ S1 is not well understood.

Average lifetime of internal conversion process is 10-12sec or less.

Fluorescence

This process of emitting a photon for deexcitation of S1 to S0.

Average lifetime of fluorescence is 10-9~10-7sec.

Intersystem Crossing (ISC)

This process of non-radiative transfer from the singlet to the triplet state.

Intersystem crossing are most common in molecules that contain heavy atoms, such as iodine or bromine.

Interaction between the spin and orbital motions becomes large in the presence of such atoms.

Phosphorescence

From T1, the molecule can return to S0 by emission of photon.

A triplet-singlet transition is much less probability than a singlet-singlet conversion.

The excited triplet state with respect to emission ranges from 10-4 to 10sec.

Rate of Deactivation Processes

Absorption- 10-14~10-15sec Vibrational Relaxation and Internal Conversion - 10-12sec or less

Fluorescence- 10-9~10-7sec Phosphorescence- 10-3~10sec

Transition Types in Fluorescence

Fluorescence process seldom results from absorption of UV radiation of wavelengths lower than 250nm because such radiation is sufficiently energetic to cause deactivation of the excited states by pre-dissociation or dissociation

200nm radiation corresponds to about 140Kcal/mol; most molecules have at least some bonds that cane be ruptured.

Quantum Efficiency Comparisonfor →* and n→* transition The majority radiation of fluorescent compoun

ds is produced by a transition involving either the n→* or the →* excited state.

The molar absorptivity of a →* transition is ordinarily 100 to 1000 times greater than for an n→* process.

The inherent lifetime associated with a →* transition is shorter (10-7 to 10-9 sec compared with 10-5 to 10-7 sec for an n→* process) and kf is larger. The rate constant for intersystem crossing ki is smaller for →* excited states.

Fluorescence is more commonly associated with →* states than with n→* states because the deactivation processes that compete with fluorescence are less likely to occur.

Fluorescence and Structure

The most intense and most useful fluorescent behavior is found in compounds containing aromatic functional groups with low energy →* transition levels.

Compounds containing aliphatic and alicyclic carbonyl structures or highly conjugated double bond structure may also exhibit fluorescence.

H3C CH3

CH3

CH2OH

CH3 CH3 CH3

Organic Compounds for Fluorescence

Vitamin A has a blue fluorescence with an emission maximum at approximately 500nm in ethanol.

highly conjugated compound

Fluorescence Instrumentation

Source- xenon-arc lamp for higher wavelengths and mercury-arc lamp for shorter wavelengths

Lensing system- quartz lenses can be used in the ultraviolet wavelength range (200-380nm), and glass lenses can be used in the visible wavelength range (380nm-700nm)

Fluorescence Instrumentation

Emission wavelength selector: this system is generally placed at an angle of 90o with respect to the excitation axis to minimize interferences from transmitted and scattered exciting light.

Beam Splitter: A beam splitter is a mirror of partial reflectance. A proportion of light would be reflected from the beam splitter while the rest would be transmitted through it unaffected.