LABMAN MODEL LMMAS-100

MICROBIAL AIR SAMPLER

Model LMMAS-100 Microbial Air Sampler is only a

sampling device, which is Sieve Impact System

based on Andarson Impaction Principle, which is

widely accepted and proven world wide.

Air impaction onto Culture Media of Petri dish ( 90

mm ) taking Micro-organism on the media and

growing colonies are counted after proper incubation.

LMMAS-100 comply with the guidelines as specified

in the guidelines as specified in ISO 14698-1

Model LMMAS-100 Microbial Air Sampler

Perforated Grid ( 488 holes )

Petri – Dish ( Ø 90 mm )

COLLECTION EFFICIENCY

A Radial Fan , made of Antistatic Plastic Resin controlled by Flow Sensor, accurately regulates Air Flow Real Time at 100 litrs/ minutes

Validation work according to BS EN ISO 14698-1:2003 Annex B has shown the Air Sampler to have Physical efficiency of 95% for particle, 0.8 to 19.0 micron in size

SPECIFICATIONS Nominal Air Flow Rate : 100 litres of Air per minute

Impact Speed : Approx. 20m/sec.

Air Sample Volumes : 10 – 6000 litres; upto 100

volumes can be selected and stored in the instrument

(User mode ). The maximum volume of Air per cycle is

6000.

Sampling Head : High quality aviation aluminum

Fan : Antistatic Plastic Resin

Housing : High quality aviation aluminum

Battery Pack : Rechargeable Ni/MH (without memory

effect) . Battery will last for 20 hrs (age dependent) after

charging of 6 hrs.

Keypad & Display Unit

APPLICATIONS

- Microbiology Labs

- Pathology Labs

- Biotechnology Labs

- Hospitals ( Operation theatre Complex )

- Clean room set up of Vaccine Units

……. many more applications