Analysis of Beverages using Large Volume Static Headspace / GCMS
hit “esc” to quit hit any key to continue
Presentation Overview• Introduction to Large Volume Static
Headspace (LVSH) Technique
• Advantages of LVSH/GC(MS)
• Analytical Procedure
• Examples of Beverages Analyzed
• Remote Headspace Analysis
• Summary
hit “esc” to quit hit any key to continue
GC/MS Detection Limits vs Headspace Sample Volume
0.001 0.01 0.1 1.0 10 100 1000
1000
100
10
1
0.1
0.01
10-3
10-4
10-5
PPM
LOD
MINIMUM VOLUME REQUIRED (CC)
Vol vs Conc.
Loop Preconcentration
hit “esc” to quit hit any key to continue
Large Volume Headspace Requires Preconcentration
HP 5973 GCMS Volume Reduction Large Volume
Inlet or Autosampler
hit “esc” to quit hit any key to continue
Large Volume Static Headspace Autosampler
• Introduces 10-200cc of headspace to the 7100 Preconcentrator
• Accommodates solids and liquids
• Detection limits 100x lower than conventional loop injection headspace
• “See what you can smell”
hit “esc” to quit hit any key to continue
LVSH Sample Platforms and Enclosures
• Samples are placed in disposable vials, reducing potential for contamination of sample platforms.
• Sample Platforms use quick connects, simplifying removal for oven bakeout.
hit “esc” to quit hit any key to continue
Advantages of Large Volume Static Headspace (LVSH)
• Cleaner. No aerosols or foaming to contaminate sample transfer lines
• Larger Sample Loading• Low detection limits without sample heating.
– Less chance of producing artifacts.– Sample remains in natural state.
• Ideal for kinetic / aging studies– Outgassing rates– Effects of Atmosphere and Storage Temperature
• Better recovery of reactive compounds– Strong carbon based adsorbents are avoided– Inert, Silonite tubing used throughout
• No sample fractionation (SPME)
hit “esc” to quit hit any key to continue
Food and Beverage Analysis by Large Volume Static Headspace (LVSH) / GCMS Analysis
HP 5973 GCMS 7100 7032L 4600
7100 3-Stage Preconcentrator
7032L 21-Position LVSH/MiniCan Autosampler
4600 Multi-Channel Standards Diluter
hit “esc” to quit hit any key to continue
7032-LVSH / 7100 /5973 Large Volume Static Headspace Analyzer
GCMS
6
7100 7032-LVSH
SL I/O SL I/O
M1
M2
M3
hit “esc” to quit hit any key to continue
PCB Integrated wiring for improved reliability
Modular design.
Easy replacement of traps
MFC downstream of traps
7100 Open Architecture
hit “esc” to quit hit any key to continue
• Silonite tubing used throughout.
•Complete heating of sample flow path. Easily accessed for trap and tubing replacement.
•External Heaters outside each cryotrap
7100 Heated Flow Path
7100 Preconcentrator
hit “esc” to quit hit any key to continue
SiloniteTM Fused Silica Lined Tubing
hit “esc” to quit hit any key to continue
Silonite Tubing VOC Adsorption/Absorption Study
• Less adsorption of Diethylbenzene on Silonite tubing. No deactivation needed.
• Faster sample equilibration with tubing surface.
• Less potential for carryover.
50
100
150
200
250
300
350
1 2 3 4
Silonite
SilcoSteel
Deact.SilcoSteel
R
E
S
P
O
N
S
E
Flush Time (Min.)
Silonite
Brand B
Deact. Brand B
hit “esc” to quit hit any key to continue
InternalStandard
CalibrationStandard
Glass Beads Tenax
Cryogen in
CO2
Focuser
ToGC
VOCs
7100 Water and CO2 Management Techniques
H2O
Helium
PUMPMFC
Helium Carrier
Removing Water and CO2
Sample
hit “esc” to quit hit any key to continue
Foods,Packaging
MP&T Glass Beads (-150C)
Tenax (-30C)
High CO2Beverages
ECTD Empty (-150C)
Tenax (-50C)
High CO2/EtOHBeverages(Heavy-Ends)
MP&T Tenax (20C) Tenax (-30C)
High CO2/EtOHBeverages(Light-Ends)
ECTD Tenax (-10C) Tenax (-60C)
3-Stage Flexibility for GCMS Analysis of Multiple Sample Types and Matrices
Sample Type Mode Trap1 Trap2
Trap 3 - On-column focuser for capillary GC/MS
hit “esc” to quit hit any key to continue
Improper Water Management
400cc VOC STD, Full Scan 15-270 amu
hit “esc” to quit hit any key to continue
Vinyl Chloride
y = 285937x - 13186
R2 = 1
-500000
500000
1500000
2500000
3500000
4500000
5500000
6500000
7500000
0 5 10 15 20 25
ppbv
MS
D A
bu
nd
an
ce Vinyl Chloride 100x Calibration Curve usingVariable Volume / Single Standard Calibration
hit “esc” to quit hit any key to continue
3 .5 0 3 .55 3 .6 0 3 . 6 5 3 .70 3 .7 5 3 . 80 3 .5 8 3 . 9 0 3 . 95
7 5 0 0 02 6 0 0 0
2 5 0 0 0
2 4 0 0 0
2 3 0 0 0
2 2 0 0 0
2 1 0 0 0
2 0 0 0 0
1 9 0 0 0
1 8 0 0 0
1 7 0 0 0
1 6 0 0 0
1 5 0 0 0
1 4 0 0 0
1 3 0 0 0
1 2 0 0 0
1 1 0 0 0
1 0 0 0 0
9 0 0 0
8 0 0 0
7 0 0 0
6 0 0 0
5 0 0 0
4 0 0 0
3 0 0 0
2 0 00
1 0 0 0
0
7 0 0 0 0
4 5 0 0 0
6 5 0 0 0
5 0 0 0 0
4 0 0 0 0
3 0 0 0 0
2 5 0 0 0
3 5 0 0 0
2 0 0 0 0
1 5 0 0 0
1 0 0 0 0
5 0 0 0
6 0 0 0 0
5 5 0 0 0
E than o lm /z 4 54 .35 m in0 .17 ppb .
M ethan o lm /z 3 13 .67 m in .0 .49 ppb
Fast Injection and Proper Water Management Minimizes Tailing of Polar VOCs
Column: HP1, 60m, 0.32mm ID, 1um film.
Flow rate: 1.5 ccm
Carrier: He
hit “esc” to quit hit any key to continue
Advantages of 7100 Multi-Stage Preconcentration
• Multiple Preconcentration Options Depending on Target Compounds and Matrix (CO2, ETOH)
• Reduced Temperature Trapping Minimizes Sample Stress, Maximizing Recovery of Thermally Labile Flavor and Odor Compounds
• Silonite Tubing Creates an Extremely Inert Sample Flow Path to Further Improve Recovery
• Demonstrated Recovery of Sulfur and Nitrogen Compounds
hit “esc” to quit hit any key to continue
LVSH Analysis of Yuban Coffee - 100cc
hit “esc” to quit hit any key to continue
100cc LVSH - Coke/Pepsi Challenge
Coke
Pepsi
hit “esc” to quit hit any key to continue
TreeTop Apple Juice100cc, LVSH60m, HP1, 0.32mm ID, 1um7032LVSH/7100/5973
Apple Juice by LVSH / GCMS
hit “esc” to quit hit any key to continue
Beer by LVSH/GCMS
Beer, ETOH Removed200cc, LVSH60m, HP1, 0.32mm ID, 1um7032LVSH/7100/5973
hit “esc” to quit hit any key to continue
Beer by LVSH/GCMSBeer, ETOH RemovedFront End, Post ETOH200cc, LVSH60m, HP1, 0.32mm ID, 1um7032LVSH/7100/5973
1
2
3
4
1 Acetone2. Ethyl Formate3. Dimethyl Sulfide4. Propanol
hit “esc” to quit hit any key to continue
Beer Light Ends by LVSH/GCMS
Beer, ETOH M1 RetainedCold Trap DehydrationLight Ends200cc, LVSH60m, HP1, 0.32mm ID, 1um7032LVSH/7100/5973
hit “esc” to quit hit any key to continue
Beer Light Ends by LVSH/GCMS
Beer, ETOH M1 RetainedCold Trap DehydrationLight Ends200cc, LVSH60m, HP1, 0.32mm ID, 1um7032LVSH/7100/5973
Acetaldehyde
COS
hit “esc” to quit hit any key to continue
2% Milk by LVSH/GCMS
2% Milk100cc, LVSH60m, HP1, 0.32mm ID, 1um7032LVSH/7100/5973
hit “esc” to quit hit any key to continue
2% Milk100cc, LVSH60m, HP1, 0.32mm ID, 1um7032LVSH/7100/5973
Dimethyl Sulfide
2% Milk by LVSH/GCMS
hit “esc” to quit hit any key to continue
MiniCan Samplers Allow Direct Sampling of Gas Phase Flavors and Odor Compounds
• Collection of Remote Headspace Samples Improves Monitoring Effeciency.
• MiniCan Samplers are SiloniteTM Coated to reduce surface losses.
• Monitor Process Streams, Holding Tanks, Reactors, etc.
hit “esc” to quit hit any key to continue
Four Models of MiniCans
MC400 MC400L MC400S MC400V
hit “esc” to quit hit any key to continue
7032 Autosampler
Gas Phase Analysis Using MiniCan Silonite Coated Samplers
Surrogate Spiking
hit “esc” to quit hit any key to continue
5973/7100/7032 200cc Injection10 PPB Sulfur Std., ECTD
1
2
3 4
6
5
1. CO2
2. H2S
3. COS
4. MeSH
5. DMS
6. CS2
hit “esc” to quit hit any key to continue
Si l coCAN
0
20
40
60
80
100
120
140
160
0 20 40 60 80 100
Rel at i ve Humi di ty (%)
Recovery (%)
Ethyl mercaptan
Propyl mercaptan
s- butyl mercaptan
t- Butyl mercaptan
SUMMA CANI STER
0
20
40
60
80
100
120
140
160
0 20 40 60 80 100
Rel at i ve Humi di ty (%)
Recovery (%)
Ethyl mercaptan
Propyl mercaptan
s- butyl mercaptan
t- Butyl mercaptan
2-Day Stability Study of Sulfur Compounds in
Fused Silica Lined vs Electropolished Canisters
7000 / HP5973 Data
hit “esc” to quit hit any key to continue
5-PointSulfur Gas
Calibrations
H2S
MeSH CS2
DMS
hit “esc” to quit hit any key to continue
Support for R&D and QA
• Fully Characterize Products down to Olfactory Detection Limits
• Identify Good and Bad “Markers” in product or packaging
• Monitoring on-line (Sequence Looping)
• Monitoring at-line (Sampling Ports)– Solids, Liquids by LVSH– Gases by MiniCan Samplers
hit “esc” to quit hit any key to continue
Conclusion• Large Volume Static Headspace (LVSH) is a new analytical
tool available to food and beverage Chemists which enhances the detection of headspace components using GC and GCMS analysis.
• Maintaining the sample in a more natural state allows the analysis of “normally occurring” odor and flavor compounds.
• Inert flow paths and reduced temperature trapping increases the range of analytes that can be recovered, including oxygen, nitrogen, and sulfur containing compounds.