History of the Prothrombin Time and Development of the INR

Post on 16-Jan-2016

60 views 0 download

Tags:

description

European Concerted Action on Anticoagulation. History of the Prothrombin Time and Development of the INR. Leon Poller. Mean results with different thromboplastins in 30 patients. Poller L. Acta Haemat. 32:292-298 (1964). Mean doses of warfarin from different geographical locations. - PowerPoint PPT Presentation

Transcript of History of the Prothrombin Time and Development of the INR

History of the Prothrombin Time and Development of the INR

Leon Poller

European Concerted Action on Anticoagulation

Mean results with different thromboplastins in 30 patients

Poller L. Acta Haemat. 32:292-298 (1964)

Mean doses of warfarin from different geographical locations.

Striped boxes indicate rabbit reagent.

Bleeding events with Simplastin (ISI 2.5) and Manchester Reagent (ISI 1.0) control.

Hull et al (1982). Prothrombin Ratio 2.0 Simplastin = 4.7 INR.

Figure 1 Figure 2

ACCP Consensus Meeting

An example of the orthogonal regression line

Calibration lines from the seventeen participant centres

The hierarchical structure of thromboplastin calibration

IRP = International Reference Preparation. *Now discontinued.

IRR = International Reference Reagent. IS = International Standard.

International reference preparations for thromboplastins 2003

INR = Prothrombin Ratio (PR)ISI

Causes of Erroneous INRs (1)

• Pretest variablesErroneous prothrombin ratio determination

Variations in manual techniqueCoagulometer effects

• Incorrect normal value due to: Failure to use MNPT Error in MNPT due to: (a)   unrepresentative selection; (b)   technical faults; (c)   non-use of geometric mean.

Causes of Erroneous INRs (2)

• Incorrect ISI calibration • Drift of ISI • Poor distribution of coumarin test samples,

inadequate numbers of test samples• Incorrect choice of IRP • Incorrect transformation of PR to INR

• ‘Rogue’ plasmas (outliers)

Errors in PT Determination • Pretest variables • blood collection • citrate • container • storage time • temperature • evacuated and non-evaculated tubes • variations in manual technique • coagulometers The overall trend of coagulometers is to accelerate the normal PT

with the PR distorted.System-related ISI of limited value.

Coagulometer Effects of ACL and MLA

The ISI of thromboplastins and effect of ISI on the width of therapeutic range in prothrombin ratios.

Percentage deviation of mean INR from the IRP value with various thromboplastin reagents during the

induction phase

Reagent C

Reagent B

Reagent A

Reagent D

Reliability of monitor INR on CoaguChek Mini and TAS PT-NC at 10

centres

No. of samples

Displayed INR

“true” INR

% difference

Mean displayed

INR difference

TAS PT-NC 536 2.91 2.52 +15.2

21.3

CoaguChek Mini

536 2.40 2.59 -7.1

Br Med J 2003;327:30-34

DAILY TELEGRAPH 23 SEPTEMBER 2003

Proportion of time in INR ranges

*INR results for all weeks, including first 3 weeks

Lancet 1998;352:1505-9