Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic...

Post on 09-Jul-2020

0 views 0 download

Transcript of Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic...

10/27/14

Atomic Force Microscope Vecco Dimension 3100

(supported by Bruker)

Location White Bldg.

Room 613

10/27/14 Atomic Force Microscope

AFM Control unit and software Atomic Force Microscope

10/27/14 Atomic Force Microscope Components

(S)canning (P)robe (M)icroscope head Stage holder and (2) types of tips 1. Tapping 2. Contact 3. Magnetic (not shown)

10/27/14

Atomic Force Microscope Tips

Tapping tip on holder.

Zoom in on tip.

www.brukerafmprobes.com

Typical Silicon Tips 5-20 nm tip radii 20-100 N/m stiffness

10/27/14 Atomic Force Microscope Tips

Contact tip on holder.

Zoom in on tip.

www.brukerafmprobes.com

Important to match the stiffness of the tip to the sample being measured.

10/27/14 Atomic Force Microscope Simplistic Operation Click to play

10/27/14 Example of AFM measurements – Tapping Mode

Nanoindentation – Optical image

Nitinol material

10/27/14 AFM scan - 30 um x 30 um

Nitinol material – Limitations of scan 100 um x 4.8 um deep

10/27/14 AFM scan - 30 um x 30 um – isometric

Nitinol material

10/27/14 AFM Geometry Analysis

Nitinol material

10/27/14 AFM Geometry Analysis – Vertical Distance

Nitinol material

10/27/14 AFM Geometry Analysis – Rmax

Nitinol material

10/27/14 AFM Geometry Analysis – Depth of indent

Nitinol material

9/9/14 AFM Geometry Analysis – Angle of indent

Nitinol material

10/27/14 Picking the right AFM Tip – Size matters

Click to play

10/27/14

Understanding contamination and its effects Click to play

• In ambient air, surface are covered with contamination • A probe encounters contamination when approaching a surface • Capillary forces pull the contamination up onto the probe

Contact AFM – Force Microscopy

10/27/14 Using contact mode with Force modulation

Consider a simple case of Tip / Sample interaction

-700

-600

-500

-400

-300

-200

-100

0

100

200

300

0 500 1000 1500 2000 2500

Def

lect

ion

(nm

)

Extension (nm)

Extend

Retract

The X & Y of the scanner is fixed. It moves in Z as shown by Extension.

Rubber sample material

A

Van der Waals forces can be seen as the tip approaches the surface (A).

10/27/14 Tip / Sample interaction

-700

-600

-500

-400

-300

-200

-100

0

100

200

300

0 500 1000 1500 2000 2500

Def

lect

ion

(nm

)

Extension (nm)

Extend

Retract

These attractive forces pull the tip down.

Rubber sample material

B Point of contact (B).

A

10/27/14 Tip / Sample interaction

-700

-600

-500

-400

-300

-200

-100

0

100

200

300

0 500 1000 1500 2000 2500

Def

lect

ion

(nm

)

Extension (nm)

Extend

Retract

Extension continues and the cantilever bends upward.

Rubber sample material

B

C

Until (C), where the extension stops and so does the deflection.

Point of contact (B).

10/27/14 Tip / Sample interaction

-700

-600

-500

-400

-300

-200

-100

0

100

200

300

0 500 1000 1500 2000 2500

Def

lect

ion

(nm

)

Extension (nm)

Extend

Retract

C – As the scanner retracts, the deflection decreases (C to D) – red curve. The cantilever relaxes.

Rubber sample material

D

C

There may be surface attraction between the tip and sample which keeps the tip in contact with the surface. A monolayer of water can also provide this attraction.

10/27/14 Tip / Sample interaction

-700

-600

-500

-400

-300

-200

-100

0

100

200

300

0 500 1000 1500 2000 2500

Def

lect

ion

(nm

)

Extension (nm)

Extend

Retract

E – As this force is overcome, the tip snaps quickly back as shown from by the vertical trajectory from D to E.

Rubber sample material

D

E

10/27/14 Tip / Sample interaction

-700

-600

-500

-400

-300

-200

-100

0

100

200

300

0 500 1000 1500 2000 2500

Def

lect

ion

(nm

)

Extension (nm)

Extend

Retract

Force Curve Summary

Rubber sample material

D

E

A

B

C

Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the Bruker Dimension 3100 atomic force microscope. Two scans are made over the same area.

The first scan determines the topology of the surface.

The second scan is done at a fixed height (200 nm) above the surface. Any changes in the phase of resonance of the tip are due to the magnetic field of the specimen.

10/27/14

MFM Cross-Sectional Image of a Carburized Duplex 2205 Stainless Steel

surface

g 20 mm

Magnetic Ferrite

Nonmagnetic Austenite: the carbon at the surface prevents polishing-induced martensite

Austenite with some polishing-induced Surface Martensite (which is magnetic)

10/27/14

10/27/14

Thank you