2100 System Signal Conditioning Amplifier - vishaypg.com · For technical questions, contact...

Post on 12-Apr-2018

219 views 0 download

Transcript of 2100 System Signal Conditioning Amplifier - vishaypg.com · For technical questions, contact...

For technical questions, contactmicro-measurements@vishaypg.com

Micro-Measurements

Document No.: 11253Revision: 01-Oct-2012

www.micro-measurements.com1

2100 System

CONFIGURATION

A 2100 System consists of:

•Onetofivemodules—Model2120BStrainGageConditioner/Amplifier(twochannels/module)

•OneModel2110BPower Supply

•OneModel2150RackAdapter

OR

•Oneortwomodules—Model2120BStrainGageConditioner/Amplifier(twochannels/module)

•OneModel2110BPowerSupply•OneModel2160BPortableFour-ChannelEnclosure

ADDITIONAL DETAILS

•Aseparate bridge power switchremovesbridgeexcitation,excitation,enablingtheoperatortodetectunwantedsignalsduetoelectricalinterferenceand/ornoise,thermocoupleeffects,andshiftsoftheinstrumentzeroduringalong-termtest.Thisfeatureisanabsolutemustfordynamictesting,andforvalidatingtest results.

•Anadjustable bridge excitation control on each channelpermitsexcitationtobesetasspecifiedbythestraingageortransducermanufacturer.Italsoallowsforanyspecialconsiderationwhichmaybedictatedbythetestmaterial;forexample,thepoorthermalconductivitynormallyassociatedwithplastics.

• Inadditiontoadjustablebridgeexcitation,eachchannelhas its own regulator circuit.Thispreventsinteractionofadjacentchannelsduringsetuporoperation.

•Eachchannelhasacontinuously variable gain control.Incombinationwithrecommendedexcitation,theindependentgaincontrolcanprovidealargeoutputsignalsothatsmallsignalscanberesolvedwithoutoverpoweringthestraingageortransducer.

•AnLED display for each channel gives positive indicationofamplifierandresistivebalance.Thiscapabilityacceleratessetupandverifiestension/compressionloading.

•Easilyreadreference marks on the setup meter indicateacceptablelinevoltageandproperoperationofinternal power supplies.

•AswitchcontainedintheModel2110BPowerSupplyallowsadjustmentwhentheline voltage is too high or too low.

•The2100Systemprovidestrue quarter-bridge, three-leadwire capability,includinginternaldummiesandsufficientplugconnectionsforremoteshuntcalibration.

•AconvenientnetworkintheModel2120BStrainGageConditioner/Amplifierallows the operator to change the factory-supplied shunt values, as well as shunt anyarmofthebridge,asrequired.

FEATURES

•Acceptsfull,half,orquarterbridges;allbridge-completiongagesbuiltin,including120/1000-and350-ohmdummies

•Fullyadjustableandregulatedbridgeexcitationoneachchannel;upto12VDCbyfront-panelcontrol

•Continuouslyvariableamplifiergainupto2100byfront-panel control

•Separatebridge-powerswitch•Output10VDCat100mA,short-circuit-proofandcurrentlimitingstandard

•LEDnullindicatorsprovidedoneachchanneltoindicateamplifierandbridge-balancecondition

•Highstabilitywithtemperatureandtime•Frequencyresponseupto50kHz•Directchannel-by-channeldisplayofdata,withoptionalpeakhold/retentioncapability

DESCRIPTION

Thedemandsoftoday’smeasurementapplicationsaremoreexactingthaneverbefore.Aninstrumentationsystemmustprovidedurabilityandversatility,reliabilitywitheaseofoperation,andeconomywithnosacrificeofaccuracy.

The2100Systemwasengineeredwithalloftheserequirementsinmind,andtoprovideadurable,multi-channelsignalconditioner/amplifiersystemcapableofperformingequallywellinawidevarietyoftestapplicationsandenvironments.Andthe2100Systemhas proven itself through applications ranging from measurements on the ocean floor to testing of the space shuttle.

The2100Systemacceptslow-levelsignals,andconditionsandamplifiesthemintohigh-leveloutputssuitableformultiple-channelsimultaneousdynamicrecording.The2100Systemiscompatiblewithstripcharts,magnetictapeandX-Yrecorders.

Straingage,load/pressuretransducerandnickeltemperaturesensorinputscanbehandledbythe2100 System without any rewiring.

Animportantdesignobjectiveachievedisminiaturizationofthesystemwhilemaintainingadequatespacingaroundthe front-panel controls. All operational controls are locatedonthefrontpanelformaximumsetupefficiency.Frequentlyusedcontrolsarefinger-operated,whileinitialsetupadjustmentsaremadethroughthefrontpanelwithascrewdriver.

Continuouslyvariableamplifiergainisachievedviaalockingten-turnconcentric-dialcountingknob,whichpermitsresettingtoapredeterminedvalueforrepeatingroutine tests.

Acombinationofintegratedcircuitsanddiscretecomponentsassuresmaximumperformanceandeaseofserviceatthelowestpossibleprice.

Signal Conditioning Amplifier

Document No.: 11253Revision: 01-Oct-2012

For technical questions, contact micro-measurements@vishaypg.com

Micro-Measurements

www.micro-measurements.com2

2100 System

Document No.: 11253Revision: 01-Oct-2012

SignalConditioningAmplifier

MODEL 2120B STRAIN GAGE CONDITIONER AMPLIFIERAtwo-channelplug-inamplifiermodulethatincludesbridgecompletion,bridgebalance,amplifierbalance,bridgeexcitationregulator,andshuntcalibration.

LED DISPLAY Setup/Indicator for amplifierbalance, bridge balance, tension/compression

BRIDGE BALANCEResistively balances the bridge;standard locking knob; digital locking knob (”K” option)

GAIN RANGE AND VERNIERVaries amplifier gain between1–2100

BRIDGE EXCITATIONVaries bridge excitationbetween 0.5–12 VDC

AMPLIFIER BALANCEAdjusts amplifier offset

SHUNT CALIBRATION(2 points)

BRIDGE EXCITATION (on/off)Removes bridge excitation

Front Panel

STRAIN GAGE/TRANSDUCER INPUTQuarter, half, and fullbridge

OUTPUT ±10 VDC, ±100 mA

POWER RECEPTACLE

Rear Panel

SPECIAL PORTION OF PRINTED CIRCUIT BOARDFOR SHUNT CALIBRATION RESISTORS AND JUMPERS

SPECIFICATIONS

Allspecificationsinthisdatasheetarenominalortypicalat+23°Cunlessnoted.Performancemaybedegradedinthepresenceofhigh-levelelectromagneticfields.

InputsQuarter(120Ω/1000Ωand350Ω),halfandfullbridge (50-1000Ω).Quarter-bridgedummygagesprovided.

Bridge Excitation0.5to12VDC(adjustableforeachchannel)with120Ω full-bridgeload.Short-circuit current: <40 mARipple,noise,and10%linechange:±2mVmax.Loadregulation:±0.2%no-loadto120Ωload(10%linechange)

Bridge Balance±2000με(quarter,half,or350Ωfullbridge),rangecanbechangedbyinternaljumperto±4000μεor±6000με

CalibrationTwo-position(centeroff)toggleswitchStandardfactory-installedresistors(±0.1%)simulate ±1000μεatGF=2

Amp Gain1to2100continuouslyadjustable±1%.

BandpassDCto5kHz(min):-0.5dB(-5%)DCto15kHz:-3dBCanbeextendedbyinternaljumperto:DCto17kHz:-0.5dBDCto50kHz:-3dB

Amp Input

Temperature coefficient of zero ±1μV/°CRTI*,±210μV/°CRTO** -10°Cto+60°C(after30minutewarm-up)

Noise RTI: (350Ω source impedance) 1μVp-pat0.1Hzto10Hz 2μVp-pat0.1Hzto100Hz 2μVRMSat0.1Hzto50kHz

*Referredtoinput **Referredtooutput

For technical questions, contactmicro-measurements@vishaypg.com

Micro-Measurements

Document No.: 11253Revision: 01-Oct-2012

www.micro-measurements.com3

2100 System

SignalConditioningAmplifier

Noise RTO 50μVp-pat0.1Hzto10Hz 80μVp-pat0.1Hzto100Hz 100μVRMSat0.1Hzto15kHz 200μVRMSat0.1Hzto50kHz

Input Impedance >100MΩ(balancelimitresistor disconnected)

Common-Mode Rejection (DCto60Hz)

GainMultiplier CMR(dB)

X2 67

X20 87

X200 100

Source Current ±10nAtypical;±40nAmax.

Ouput±10V(min)at±100mA Current limit: 140 mA

Size5.25Hx2.94Wx10.97Din(133x75x279mm)

Weight 2.2lb(1.0kg)

MODEL 2110B POWER SUPPLY

Aplug-inmodulecapableofpoweringuptotenchannels(fiveModel2120Bmodules)atamaximumratedvoltage or current.

Providesinitialbridgeandamplifiervoltages.Allsupplies arecurrent-limitedagainstamplifiermalfunction.

BRIDGE-VOLTS METER Used to set up/monitor bridge excitation, also lineand power supply levels

CHANNEL SELECTORAC monitors ac line input.DC monitors the power supplies. Positions 1–10select and display bridgeexcitation for each channel

PILOT LAMP Indicates main power

POWER SWITCHMain power on-off

EXTERNAL METERUsed with an external digital voltmeter to precisely adjust bridge excitation

SPECIFICATIONS

Ouputs±15Vat1.2Aand+17.5Vat1.1A;allregulatorscurrent-limitedagainstoverload

Input107,115,214,230VAC±10%50/60Hz(selectedinternally)Power:40Wtypical,100Wmax.

Meter0to12VDC(withswitch)toreadbridgeexcitation.AlsoACinputandDCoutputgo/no-gomonitor

Size5.25Hx2.44Wx12.34Din(133x62x313mm)

Weight6.7lb(3.1kg)

MODEL 2150 RACK ADAPTER

AprewiredrackadapterwhichacceptsoneModel2110BanduptofiveModel2120BStrainGageConditionerAmplifiers.Ithasitsownfuseandpowercordandcanbehousedinanystandard19-in(483-mm)electronicequipment rack.

Power2-ft(0.6-m)3-wirelinecord;10-ft(3-m)extensionavailableFuse:1Asize3AG(32x6.5dia.mm)Receptacletoacceptlinecordfromadjacent2150RackAdapter

Size5.25Hx19Wx14.17Din(133x483x360mm)

Weight6.6lb(3.0kg)

LINE CORDAUXILIARY RECEPTACLE FUSE

MODEL 2150 FRONT

MODEL 2150 REAR

For technical questions, contact micro-measurements@vishaypg.com

Micro-Measurements

www.micro-measurements.com4

2100 System

Document No.: 11253Revision: 01-Oct-2012

SignalConditioningAmplifier

MODEL 2160B PORTABLE FOUR-CHANNEL ENCLOSURE

Model 2160:Aprewired,fusedenclosurewhichhousesuptothree(3)modules.Acarryinghandleensuresmaximumportability.Anadditionalsnap-downbailsupportonthebottomcanbeusedtoelevatethe2160forexcellentworkefficiencyduringbench-topoperation.TheModel2160wouldbesubstitutedfortheModel2150whentwoorfourchannelsandmaximumportabilityarerequired.

SPECIFICATIONS

Size5.55Hx8.75Wx13.80Din(141x222x350mm)

Weight5.2lb(2.4kg)

Vishay Precision Group, Inc.

www.vpgsensors.com1

Legal Disclaimer Notice

Document No.: 63999Revision: 15-Jul-2014

DisclaimerALL PRODUCTS, PRODUCT SPECIFICATIONS AND DATA ARE SUBJECT TO CHANGE WITHOUT NOTICE.

Vishay Precision Group, Inc., its affiliates, agents, and employees, and all persons acting on its or their behalf (collectively, “VPG”), disclaim any and all liability for any errors, inaccuracies or incompleteness contained herein or in any other disclosure relating to any product.

The product specifications do not expand or otherwise modify VPG’s terms and conditions of purchase, including but not limited to, the warranty expressed therein.

VPG makes no warranty, representation or guarantee other than as set forth in the terms and conditions of purchase. To the maximum extent permitted by applicable law, VPG disclaims (i) any and all liability arising out of the application or use of any product, (ii) any and all liability, including without limitation special, consequential or incidental damages, and (iii) any and all implied warranties, including warranties of fitness for particular purpose, non-infringement and merchantability.

Information provided in datasheets and/or specifications may vary from actual results in different applications and performance may vary over time. Statements regarding the suitability of products for certain types of applications are based on VPG’s knowledge of typical requirements that are often placed on VPG products. It is the customer’s responsibility to validate that a particular product with the properties described in the product specification is suitable for use in a particular application. You should ensure you have the current version of the relevant information by contacting VPG prior to performing installation or use of the product, such as on our website at vpgsensors.com.

No license, express, implied, or otherwise, to any intellectual property rights is granted by this document, or by any conduct of VPG.

The products shown herein are not designed for use in life-saving or life-sustaining applications unless otherwise expressly indicated. Customers using or selling VPG products not expressly indicated for use in such applications do so entirely at their own risk and agree to fully indemnify VPG for any damages arising or resulting from such use or sale. Please contact authorized VPG personnel to obtain written terms and conditions regarding products designed for such applications.

Product names and markings noted herein may be trademarks of their respective owners.

Copyright Vishay Precision Group, Inc., 2014. All rights reserved.

Disclaimer

Legal Disclaimer Notice